Servicing
Solenoid Valves Removal
7-52
42iQTL Instruction Manual
Thermo Scientific
Use the following procedure to replace the mode valve, and optional
converter valve and prereactor valve.
Equipment Required:
Phillips drive, #1
CAUTION
Allow the oven to cool down prior to servicing.
▲
1.
Turn the instrument OFF, unplug the power cord, and remove the
cover (
Figure 2–1
).
2.
Unplug mode valve (J5), converter valve (J6), and prereactor valve (J8)
from step board 2 only.
Figure 7–53.
Conv-Mode-Preact Electrical Connection
Solenoid Valves
Removal