![Thermo Scientific 42iQTL Скачать руководство пользователя страница 108](http://html.mh-extra.com/html/thermo-scientific/42iqtl/42iqtl_instruction-manual_1105146108.webp)
Operation
Settings
3-82
42iQTL Instruction Manual
Thermo Scientific
The Gas Units screen defines how the NO, NO
2
, and NO
x
concentration
readings are expressed. The μg/m
3
, mg/m
3
, and g/m
3
gas concentration
modes are calculated using a standard pressure of 760 mmHg and a
standard temperature of 0 °C. The temperature can be changed in the
compensation screen.
Use the
and
buttons to select.
Home Screen>Settings>Measurement Settings>Gas Units
The Gas Units screen contains the following information:
●
ppb:
parts per billion.
●
ppm:
parts per million.
●
%:
percent.
●
μg/m3:
micrograms per meter cubed.
●
mg/m3:
milligrams per meter cubed.
●
g/m3:
grams per meter cubed.
Gas Units
Содержание 42iQTL
Страница 1: ...42iQTL Instruction Manual Trace Level NO NO2 NOx Analyzer 119019 00 1April2020 ...
Страница 2: ......
Страница 26: ......
Страница 34: ...Operation Instrument Display 3 8 42iQTL Instruction Manual Thermo Scientific ...
Страница 162: ......
Страница 190: ......
Страница 196: ...Maintenance Pump Rebuilding 5 6 42iQTL Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 212: ...Servicing Fan Replacement 7 8 42iQTL Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 229: ...Servicing Power Supply Replacement Thermo Scientific 42iQTL Instruction Manual 7 25 Figure 7 25 Replacing Power Supply ...
Страница 253: ...Servicing DMC Ozonator Thermo Scientific 42iQTL Instruction Manual 7 49 Figure 7 49 Remove Ammonia Scrubber ...
Страница 258: ......
Страница 292: ......
Страница 304: ......
Страница 308: ......
Страница 309: ......