![Thermo Scientific 42iQTL Скачать руководство пользователя страница 184](http://html.mh-extra.com/html/thermo-scientific/42iqtl/42iqtl_instruction-manual_1105146184.webp)
Calibration
Manual Calibration
4-22
42iQTL Instruction Manual
Thermo Scientific
Note however, that the EPA requires a minimum of one multipoint
calibration per calendar quarter. Such a quality control program is essential
to ascertain the accuracy and reliability of the air quality data collected and
to alert the user if the accuracy or reliability of the data should become
unacceptable. A compilation of this kind might include items such as dates
of calibration, atmospheric conditions, calibration factors, and other
pertinent data.
Use the following procedure to perform a zero and span check.
1.
Connect the zero gas to the SAMPLE bulkhead in a standard
instrument or to the ZERO IN bulkhead in a 42iQTL
equipped with
the zero/span and sample solenoid valve option.
2.
Allow the instrument to sample zero gas until a stable reading is
obtained on the NO, NO
2
, and NO
x
channels then record the zero
readings. Unless the zero has changed by more than ±0.010 ppm, it is
recommended that the zero not be adjusted. If an adjustment larger
than this is indicated due to a change in zero reading, a new multipoint
calibration curve should be generated.
3.
Attach a supply of known concentration of NO and NO
2
(usually
generated via an NIST traceable NO working standard and a GPT
system) to the SAMPLE bulkhead (or SPAN bulkhead for instruments
equipped with the zero/span and sample solenoid valve option) on the
rear panel.
4.
Allow the instrument to sample the calibration gas until a stable
reading is obtained on the NO, NO
2
, and NO
x
channels. If the
calibration has changed by more than
10%, a new multipoint
calibration curve should be generated.
5.
When the calibration check has been completed, record the NO, NO
2
,
and NO
x
values.
6.
Reconnect the analyzer sample line to the SAMPLE bulkhead.
The Manual Calibration screen allows the user to view and manually adjust
the zero background and span coefficient. These are used to correct the
NO, NO
2
, and NO
x
readings that the instrument generates using its own
internal calibration data.
Manual
Calibration
Содержание 42iQTL
Страница 1: ...42iQTL Instruction Manual Trace Level NO NO2 NOx Analyzer 119019 00 1April2020 ...
Страница 2: ......
Страница 26: ......
Страница 34: ...Operation Instrument Display 3 8 42iQTL Instruction Manual Thermo Scientific ...
Страница 162: ......
Страница 190: ......
Страница 196: ...Maintenance Pump Rebuilding 5 6 42iQTL Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 212: ...Servicing Fan Replacement 7 8 42iQTL Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 229: ...Servicing Power Supply Replacement Thermo Scientific 42iQTL Instruction Manual 7 25 Figure 7 25 Replacing Power Supply ...
Страница 253: ...Servicing DMC Ozonator Thermo Scientific 42iQTL Instruction Manual 7 49 Figure 7 49 Remove Ammonia Scrubber ...
Страница 258: ......
Страница 292: ......
Страница 304: ......
Страница 308: ......
Страница 309: ......