Servicing
DMC PMT Cooler and Reaction Chamber Replacement
7-34
42iQTL Instruction Manual
Thermo Scientific
Figure 7–34.
Photo Reaction Chamber Cleaning, Hardware
6.
Pull reaction chamber outwards.
Equipment Damage
Some internal components can be damaged by small
amounts of static electricity. A properly grounded antistatic wrist strap
must be worn while handling an internal component.
▲
7.
To reinstall reaction chamber, follow previous steps in reverse, making
sure to back fill the cooler with dry air or nitrogen prior to installing
reaction chamber.
8.
Reinstall the measurement bench. Refer to
Figure 7–33
and
Figure 7–
40
.