Operation
Settings
3-86
42iQTL Instruction Manual
Thermo Scientific
●
Down:
●
Manual Entry:
Shows current value. If the PMT supply cell is
pressed, user can manually change the value.
●
Target Concentration:
Displays target concentration. If target
concentration cell is pressed, user can enter target concentration.
●
Calibration Concentration:
Displays calibration concentration.
●
Auto Calibration:
When pressed, the auto-calibration process is
initiated. Please allow up to 5 minutes for calibration to complete.
PMT voltage is automatically adjusted until calibration
concentration is with 1% of target concentration.
●
End Calibration:
When pressed, the auto calibration is interrupted
and the value does not change.
●
Default PMT Supply:
When pressed, the default PMT supply value
is saved.
Содержание 42iQTL
Страница 1: ...42iQTL Instruction Manual Trace Level NO NO2 NOx Analyzer 119019 00 1April2020 ...
Страница 2: ......
Страница 26: ......
Страница 34: ...Operation Instrument Display 3 8 42iQTL Instruction Manual Thermo Scientific ...
Страница 162: ......
Страница 190: ......
Страница 196: ...Maintenance Pump Rebuilding 5 6 42iQTL Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 212: ...Servicing Fan Replacement 7 8 42iQTL Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 229: ...Servicing Power Supply Replacement Thermo Scientific 42iQTL Instruction Manual 7 25 Figure 7 25 Replacing Power Supply ...
Страница 253: ...Servicing DMC Ozonator Thermo Scientific 42iQTL Instruction Manual 7 49 Figure 7 49 Remove Ammonia Scrubber ...
Страница 258: ......
Страница 292: ......
Страница 304: ......
Страница 308: ......
Страница 309: ......