![Thermo Scientific 42iQTL Скачать руководство пользователя страница 234](http://html.mh-extra.com/html/thermo-scientific/42iqtl/42iqtl_instruction-manual_1105146234.webp)
Servicing
DMC PMT Cooler and Reaction Chamber Replacement
7-30
42iQTL Instruction Manual
Thermo Scientific
Figure 7–30.
Remove PMT Cooler Shroud
5.
Unplug J3, J1, and J6 (fan).
6.
Disconnect all plumbing depending on the options. Configuration
plumbing connections will vary.
7.
Using #2 Phillips drive, unfasten four captive hardware.
8.
Pull PMT cooler upwards.