range is active, the range mode selection status output is at 0 volts.
When the high range is active, the range mode selection status output is
at half of full-scale.
When the high range is active, the NO concentration must drop to
85% of the low NO range for the low range to become active.
In addition to each channel having two ranges, each channel has two
span coefficients. There are two span coefficients so that each range can
be calibrated separately. This is especially necessary if the two ranges are
not close to one another. For example, the low NO range is set to 0–50
ppb and the high NO range is set to 0–200 ppb.
Содержание 42iQTL
Страница 1: ...42iQTL Instruction Manual Trace Level NO NO2 NOx Analyzer 119019 00 1April2020 ...
Страница 2: ......
Страница 26: ......
Страница 34: ...Operation Instrument Display 3 8 42iQTL Instruction Manual Thermo Scientific ...
Страница 162: ......
Страница 190: ......
Страница 196: ...Maintenance Pump Rebuilding 5 6 42iQTL Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 212: ...Servicing Fan Replacement 7 8 42iQTL Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 229: ...Servicing Power Supply Replacement Thermo Scientific 42iQTL Instruction Manual 7 25 Figure 7 25 Replacing Power Supply ...
Страница 253: ...Servicing DMC Ozonator Thermo Scientific 42iQTL Instruction Manual 7 49 Figure 7 49 Remove Ammonia Scrubber ...
Страница 258: ......
Страница 292: ......
Страница 304: ......
Страница 308: ......
Страница 309: ......