![Thermo Scientific 42iQTL Скачать руководство пользователя страница 151](http://html.mh-extra.com/html/thermo-scientific/42iqtl/42iqtl_instruction-manual_1105146151.webp)
Operation
Settings
Thermo Scientific
42iQTL Instruction Manual
3-125
Use the following procedure to update firmware using the USB
connection.
1.
Copy the update*.zip file onto the USB FLASH drive using Windows
Explorer or another means. The file must be copied into the root
directory of the USB stick's file system otherwise it won't be
recognized. Also, make sure the USB stick is formatted using the FAT
or FAT32 file system.
2.
Plug a USB flash drive into the data port on the front panel of the unit
or into one of the USB ports on the side near the power cable. A pop-
up box will appear, acknowledging that a flash drive has been detected.
If a USB password has been previously set, you will be prompted to
enter the USB password to continue. Press
Enter
to continue.
3.
Press the OK button on the pop-up box to go to the Firmware Update
Via USB Drive screen.
Содержание 42iQTL
Страница 1: ...42iQTL Instruction Manual Trace Level NO NO2 NOx Analyzer 119019 00 1April2020 ...
Страница 2: ......
Страница 26: ......
Страница 34: ...Operation Instrument Display 3 8 42iQTL Instruction Manual Thermo Scientific ...
Страница 162: ......
Страница 190: ......
Страница 196: ...Maintenance Pump Rebuilding 5 6 42iQTL Instruction Manual Thermo Scientific Figure 5 2 Pump Rebuilding ...
Страница 212: ...Servicing Fan Replacement 7 8 42iQTL Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 229: ...Servicing Power Supply Replacement Thermo Scientific 42iQTL Instruction Manual 7 25 Figure 7 25 Replacing Power Supply ...
Страница 253: ...Servicing DMC Ozonator Thermo Scientific 42iQTL Instruction Manual 7 49 Figure 7 49 Remove Ammonia Scrubber ...
Страница 258: ......
Страница 292: ......
Страница 304: ......
Страница 308: ......
Страница 309: ......