Operation
Calibration
Thermo Scientific
42iQLS Instruction Manual
3-29
Home Screen> Calibration>Advanced Calibration>Manual
Calibration>Adjust Coefficients> Adjust High Range NO Span
Coefficient (dual or auto range mode)
The Adjust NO Span Coefficient screens contain the following
information:
●
Adjust NO Span Coefficient:
User manually adjusts NO span coefficient
when in single range mode.
●
Adjusted NO Concentration:
Read only. Shows adjusted NO
concentration based on adjusted NO span coefficient when in single
range mode.
●
Current NO Concentration:
Read only. Shows current NO
concentration when in single range mode.
●
Adjust High Range NO Span Coefficient:
User manually adjusts the high
range NO span coefficient when in dual or auto range mode.
●
Adjusted High Range NO Concentration:
Read only. Shows adjusted
high range NO concentration based on adjusted high range NO span
coefficient when in dual or auto range mode.
●
Current High Range NO Concentration:
Read only. Shows current high
range NO concentration when in dual or auto range mode.
●
Adjust Low Range NO Span Coefficient:
User manually adjusts the NO
low span coefficient when in dual or auto range mode.
●
Adjusted Low Range NO Concentration:
Read only. Shows adjusted NO
concentration based on adjusted low range NO span coefficient when
in dual or auto range mode.
●
Adjusted Low Range NO Concentration:
Read only. Shows current low
range NO concentration when in dual or auto range mode.
●
Calibrate:
Calibrates span coefficient by saving the newly adjusted span
coefficient.
Содержание 42iQLS
Страница 1: ...42iQLS Instruction Manual Low Source NO NO2 NOx Analyzer 117436 00 1Sep2019 ...
Страница 2: ......
Страница 18: ......
Страница 34: ...Operation Instrument Display 3 8 42iQLS Instruction Manual Thermo Scientific ...
Страница 192: ......
Страница 199: ...Maintenance Pump Rebuilding Thermo Scientific 42iQLS Instruction Manual 5 7 Figure 5 3 Pump Rebuilding ...
Страница 214: ...Servicing Fan Replacement 7 8 42iQLS Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 230: ...Servicing Power Supply Replacement 7 24 42iQLS Instruction Manual Thermo Scientific Figure 7 26 Replacing Power Supply ...
Страница 254: ...Servicing DMC Ozonator 7 48 42iQLS Instruction Manual Thermo Scientific Figure 7 50 Remove Ammonia Scrubber ...
Страница 296: ......
Страница 304: ......
Страница 316: ......
Страница 317: ......