Calibration
Calibration
Thermo Scientific
42iQLS Instruction Manual
4-13
is pressed. Pressing the Calibrate button will save the new NO
x
span
coefficient and calibrate the instrument.
3.
Enter the NO
x
calibration gas concentration using the pushbuttons,
and then press
Calibrate
to calibrate the instrument to the NO
x
calibration gas.
The exact NO
x
concentration is calculated from:
F
+
F
+
F
)
]
NO
[
+
]
[NO
(
x
F
=
]
NO
[
D
O
NO
IMP
2
STD
NO
OUT
x
Where:
[NO
x
]
OUT
= diluted NO
x
concentration at the output manifold, ppm
[NO
2
]
IMP
= concentration of NO
2
impurity in the standard NO cylinder, ppm
4.
Record the NO
x
concentration and the analyzer's NO
x
response.
Use the following procedures to prepare the NO, NO
x
, and NO
2
calibration curves.
1.
Generate several additional NO and NO
x
concentrations by decreasing
F
NO
or increasing F
D
.
2.
For each concentration generated, calculate the exact NO and NO
x
concentrations using the above equations for [NO]
OUT
and [NO
x
]
OUT
.
3.
Record the NO and NO
x
responses.
4.
Plot the analyzer responses versus the respective calculated NO and
NO
x
concentrations and draw or calculate the respective calibration
curves. For subsequent calibrations where linearity can be assumed,
these curves may be checked with a three-point calibration consisting of
a zero point, NO and NO
x
concentrations of approximately 80% of the
URL, and an intermediate concentration.
5.
Adjust the GPT system to generate a NO concentration near 90% of
the URL of the instrument range selected.
6.
Sample this NO concentration until the NO and NO
x
responses have
stabilized, then measure and record the NO concentration as [NO]
ORIG
.
Preparing NO, NO
x
, and
NO
2
Calibration Curves
Содержание 42iQLS
Страница 1: ...42iQLS Instruction Manual Low Source NO NO2 NOx Analyzer 117436 00 1Sep2019 ...
Страница 2: ......
Страница 18: ......
Страница 34: ...Operation Instrument Display 3 8 42iQLS Instruction Manual Thermo Scientific ...
Страница 192: ......
Страница 199: ...Maintenance Pump Rebuilding Thermo Scientific 42iQLS Instruction Manual 5 7 Figure 5 3 Pump Rebuilding ...
Страница 214: ...Servicing Fan Replacement 7 8 42iQLS Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 230: ...Servicing Power Supply Replacement 7 24 42iQLS Instruction Manual Thermo Scientific Figure 7 26 Replacing Power Supply ...
Страница 254: ...Servicing DMC Ozonator 7 48 42iQLS Instruction Manual Thermo Scientific Figure 7 50 Remove Ammonia Scrubber ...
Страница 296: ......
Страница 304: ......
Страница 316: ......
Страница 317: ......