Maintenance
Pump Rebuilding
Thermo Scientific
42iQLS Instruction Manual
5-5
Use the following procedure to rebuild the pump.
Equipment required:
Pump rebuild kit (qty. 2)
Phillips drive, #1 or Torque drive, T10 (depending on pump version)
Pencil or marker
Figure 5–2.
Dual Stage Pump
Note
To avoid opening the pneumatic connection between the pump
heads, service the pump as described below with the following addition:
Make steps 1 to 3 and 11 for both pump heads together.
▲
1.
Turn instrument OFF, unplug the power cord, and remove the cover.
2.
Mark the position of head parts relative to each other by drawing a line
with a pencil. This helps avoid incorrect assembly later.
3.
Undo the four screws in the head.
4.
Lift the head plate and the intermediate plate off the housing.
Pump Rebuilding
Содержание 42iQLS
Страница 1: ...42iQLS Instruction Manual Low Source NO NO2 NOx Analyzer 117436 00 1Sep2019 ...
Страница 2: ......
Страница 18: ......
Страница 34: ...Operation Instrument Display 3 8 42iQLS Instruction Manual Thermo Scientific ...
Страница 192: ......
Страница 199: ...Maintenance Pump Rebuilding Thermo Scientific 42iQLS Instruction Manual 5 7 Figure 5 3 Pump Rebuilding ...
Страница 214: ...Servicing Fan Replacement 7 8 42iQLS Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 230: ...Servicing Power Supply Replacement 7 24 42iQLS Instruction Manual Thermo Scientific Figure 7 26 Replacing Power Supply ...
Страница 254: ...Servicing DMC Ozonator 7 48 42iQLS Instruction Manual Thermo Scientific Figure 7 50 Remove Ammonia Scrubber ...
Страница 296: ......
Страница 304: ......
Страница 316: ......
Страница 317: ......