Calibration
Equipment Required
4-6
42iQLS Instruction Manual
Thermo Scientific
3.
Plot the analyzer NO response (in ppm) versus the nominal diluted
NO concentration and determine the slope, S
NOM
.
4.
Calculate the [NO] concentration of the working standard, [NO]
STD
,
from:
S
x
]
[NO
=
]
[NO
NOM
NOM
STD
5.
If the nominal NO concentration of the working standard is unknown,
generate several NO concentrations to give on-scale NO responses.
6.
Measure and record F
NO
and F
T
for each NO concentration generated.
7.
Plot the analyzer NO response versus F
NO
/F
T
and determine the slope
which gives [NO]
STD
directly. The analyzer NO
x
responses to the
generated NO concentrations reflect any NO
2
impurity in the NO
working standard.
8.
Plot the difference between the analyzer NO
x
and NO responses versus
F
NO
/F
T
. The slope of this plot is [NO
2
]
IMP
.
A source of zero air free of contaminants should be used as described earlier
in this chapter. Contaminants can cause a detectable response on the
instrument and may also react with the NO, O
3
, or NO
2
during the gas
phase titration.
Use the following definitions for the remainder of this chapter.
P
R
=
Dynamic parameter specification to ensure complete reaction of the
available O
3
, ppm-min
[NO]
RC
=
NO concentration in the reaction chamber, ppm
t
R
=
residence time of the reactant gases in the reaction chamber, min
[NO]
STD
=
Concentration of the undiluted NO standard, ppm
F
NO
=
NO flow rate, sccm
F
O
=
O
3
generator air flow rate, sccm
V
RC
=
Volume of the reaction chamber, cc
F
T
=
Analyzer demand plus 10 to 50% excess
Zero Air
Dynamic Parameter
Specifications for
Gas Titrator
Содержание 42iQLS
Страница 1: ...42iQLS Instruction Manual Low Source NO NO2 NOx Analyzer 117436 00 1Sep2019 ...
Страница 2: ......
Страница 18: ......
Страница 34: ...Operation Instrument Display 3 8 42iQLS Instruction Manual Thermo Scientific ...
Страница 192: ......
Страница 199: ...Maintenance Pump Rebuilding Thermo Scientific 42iQLS Instruction Manual 5 7 Figure 5 3 Pump Rebuilding ...
Страница 214: ...Servicing Fan Replacement 7 8 42iQLS Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 230: ...Servicing Power Supply Replacement 7 24 42iQLS Instruction Manual Thermo Scientific Figure 7 26 Replacing Power Supply ...
Страница 254: ...Servicing DMC Ozonator 7 48 42iQLS Instruction Manual Thermo Scientific Figure 7 50 Remove Ammonia Scrubber ...
Страница 296: ......
Страница 304: ......
Страница 316: ......
Страница 317: ......