Calibration
Calibration
4-14
42iQLS Instruction Manual
Thermo Scientific
7.
Adjust the O
3
generator in the GPT system to generate sufficient O
3
to
produce a decrease in the NO concentration equivalent to about 80%
of the URL of the NO
2
range. The decrease must not exceed 90% of
the NO concentration determined in Steps 5 and 6 above.
8.
When the analyzer responses stabilize, record the resultant NO
concentrations as [NO]
REM
.
9.
Press the Back button to return to the Calibration menu, and choose
Calibrate NO
2
Span Coefficient
.
The user sets the span concentration by pressing the Edit NO
2
Span
Concentration button. The Calculated NO
2
Span Coefficient button
will show what the span coefficient will be set to if the Calibrate button
is pressed. Pressing the Calibrate button will save the new NO
2
span
coefficient and calibrate the instrument.
10.
Enter the NO
2
calibration gas concentration using the pushbuttons,
and then press
Calibrate
to concentration to reflect the sum of the
following: the NO
2
concentration generated by GPT, ([NO]
ORIG
-
[NO]
REM
), and any NO
2
impurity.
The exact NO
x
concentration is calculated from:
F
+
F
+
F
]
NO
[
x
F
+
)
]
[NO
-
]
([NO
=
]
NO
[
D
O
NO
IMP
2
NO
REM
ORIG
OUT
2
Where:
[NO
2
]
OUT
= diluted NO
2
concentration at the output manifold, ppm
[NO]
ORIG
= original NO concentration, prior to addition of O
3
, ppm
[NO]
REM
= NO concentration remaining after addition of O
3
, ppm
The analyzer does a one point NO
2
span coefficient calculation,
corrects the NO
2
reading for converter inefficiency, and then adds the
corrected NO
2
to the NO signal to give a corrected NO
x
signal.
If the analyzer calculates a NO
2
span coefficient of less than 0.96, either
the entered NO
2
concentration is incorrect, the converter is not being
heated to the proper temperature, the instrument needs servicing (leak
or imbalance), or the converter needs replacement or servicing. The
NO
2
analog output will reflect the NO
2
concentration generated by
GPT, any NO
2
impurity, and the NO
2
zero offset.
Содержание 42iQLS
Страница 1: ...42iQLS Instruction Manual Low Source NO NO2 NOx Analyzer 117436 00 1Sep2019 ...
Страница 2: ......
Страница 18: ......
Страница 34: ...Operation Instrument Display 3 8 42iQLS Instruction Manual Thermo Scientific ...
Страница 192: ......
Страница 199: ...Maintenance Pump Rebuilding Thermo Scientific 42iQLS Instruction Manual 5 7 Figure 5 3 Pump Rebuilding ...
Страница 214: ...Servicing Fan Replacement 7 8 42iQLS Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 230: ...Servicing Power Supply Replacement 7 24 42iQLS Instruction Manual Thermo Scientific Figure 7 26 Replacing Power Supply ...
Страница 254: ...Servicing DMC Ozonator 7 48 42iQLS Instruction Manual Thermo Scientific Figure 7 50 Remove Ammonia Scrubber ...
Страница 296: ......
Страница 304: ......
Страница 316: ......
Страница 317: ......