Optional Equipment
Connecting External Devices
9-6
42iQLS Instruction Manual
Thermo Scientific
Table 9–4.
Analog Voltage and Current Assignment
Terminal Number
Signal Name
1
Current Out 1
2
Voltage Out 1
3 C/V
Return
1
4
Current Out 2
5
Voltage Out 2
6 C/V
Return
2
7
Current Out 3
8
Voltage Out 3
9 C/V
Return
3
10
Current Out 4
11
Voltage Out 4
12 C/V
Return
4
13
Current Out 5
14
Voltage Out 5
15 C/V
Return
5
16
Current Out 6
17
Voltage Out 6
18 C/V
Return
6
The iQ series instruments provide for the ability to calibrate the analog
outputs (both voltage and current) of the instruments. The basic procedure
for both voltage and current are the same using the following procedure:
●
Complete the connections of the recording device to the desired analog
output channel. (See page
9-5
for the channel information).
●
Calibrate the output channel low level.
Note
When calibrating the current output when using the 0-20 mA
scale, the low level will be set to 4 mA due to the inability to adjust the
actual current output to below zero.
▲
●
Calibrate the output channel full scale.
Analog Output
Calibration
Содержание 42iQLS
Страница 1: ...42iQLS Instruction Manual Low Source NO NO2 NOx Analyzer 117436 00 1Sep2019 ...
Страница 2: ......
Страница 18: ......
Страница 34: ...Operation Instrument Display 3 8 42iQLS Instruction Manual Thermo Scientific ...
Страница 192: ......
Страница 199: ...Maintenance Pump Rebuilding Thermo Scientific 42iQLS Instruction Manual 5 7 Figure 5 3 Pump Rebuilding ...
Страница 214: ...Servicing Fan Replacement 7 8 42iQLS Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 230: ...Servicing Power Supply Replacement 7 24 42iQLS Instruction Manual Thermo Scientific Figure 7 26 Replacing Power Supply ...
Страница 254: ...Servicing DMC Ozonator 7 48 42iQLS Instruction Manual Thermo Scientific Figure 7 50 Remove Ammonia Scrubber ...
Страница 296: ......
Страница 304: ......
Страница 316: ......
Страница 317: ......