Operation
Settings
Thermo Scientific
42iQLS Instruction Manual
3-95
●
Down:
●
Manual Entry:
Shows current value. If the PMT supply cell is
pressed, user can manually change the value.
●
Target Concentration:
Displays target concentration. If target
concentration cell is pressed, user can enter target concentration.
●
Calibration Concentration:
Displays calibration concentration.
●
Auto Calibration:
When pressed, the auto-calibration process is
initiated. Please allow up to 5 minutes for calibration to complete.
PMT voltage is automatically adjusted until calibration
concentration is with 1% of target concentration.
●
End Calibration:
When pressed, the auto calibration is interrupted
and the value does not change.
●
Default PMT Supply:
When pressed, the default PMT supply value
is saved.
Содержание 42iQLS
Страница 1: ...42iQLS Instruction Manual Low Source NO NO2 NOx Analyzer 117436 00 1Sep2019 ...
Страница 2: ......
Страница 18: ......
Страница 34: ...Operation Instrument Display 3 8 42iQLS Instruction Manual Thermo Scientific ...
Страница 192: ......
Страница 199: ...Maintenance Pump Rebuilding Thermo Scientific 42iQLS Instruction Manual 5 7 Figure 5 3 Pump Rebuilding ...
Страница 214: ...Servicing Fan Replacement 7 8 42iQLS Instruction Manual Thermo Scientific Figure 7 8 Replacing the Fan ...
Страница 230: ...Servicing Power Supply Replacement 7 24 42iQLS Instruction Manual Thermo Scientific Figure 7 26 Replacing Power Supply ...
Страница 254: ...Servicing DMC Ozonator 7 48 42iQLS Instruction Manual Thermo Scientific Figure 7 50 Remove Ammonia Scrubber ...
Страница 296: ......
Страница 304: ......
Страница 316: ......
Страница 317: ......