7 Maintaining the product
150
TestoSaverisPrerequisites directly under a root directory
(e.g. C:\ or D:\) so that the call-up path of the setup file is as
follows: C:\ TestoSaverisPrerequisites or
D:\TestoSaverisPrerequisites
Pos: 179 /TD/Produkt instand halten/testo Saveris/08 c Firmware System-Update durchführen Einleitung SBE @ 15\mod_1380719732363_79.docx @ 177529 @ 3 @ 1
7.9.2.
Carrying out a firmware system update
With the firmware system update, all Saveris components
in your system are updated automatically to the latest
firmware status. This ensures that all components in your
system show the current firmware version.
Pos: 180 /TD/Produkt instand halten/testo Saveris/08 c Firmware System-Update durchführen @ 12\mod_1338384658937_79.docx @ 126083 @ 35 @ 1
Important information on the firmware system update
• Please note that, with the testo Saveris data monitoring system,
two firmware system updates
(V1.X and V2.X)
are available.
First check exactly which firmware update is valid for your
system. You can check the compatibility of the firmware
versions via the order number on your Saveris base's type plate
◦
Firmware System Update V1.X
valid for Saveris systems with base article no. 0572 0
1
20,
0572 0
1
21, 0572 0160, 0572 0
1
61
◦
Firmware System-Update V2.X
valid for Saveris systems with base article no. 0572 0
2
20,
0572 0
2
21, 0572 0
2
60, 0572 0
2
61
• How long the firmware system update takes depends on the
number of Saveris components in the system. Automatic
distribution of the firmware to:
◦
Converter and extender
requires at least 3 m 2
minutes per component for the standard communication
cycle.
◦
Ethernet probe
requires at least 3 m 2 minutes per
component for the standard communication cycle.
◦
Router
requires at least 6.5 hours (all logged-in routers are
updated in parallel).
◦
Radio probe
in stationary use requires at least 16 hours (all
logged-in radio probes are updated in parallel).
• The Saveris base must be connected to the computer via USB
or Ethernet cable.
• All Saveris components must have at least firmware version
1.12 for the update. System components with firmware version
< V 1.12 can only be updated via service interface.
• For an update to version 4.3, the components must be
equipped with the following firmware version as a minimum
requirement:
Содержание 0572 0158
Страница 1: ...Measurement data monitoring with testo Saveris Small Business Edition Instruction manual...
Страница 2: ...2...
Страница 54: ...5 First steps 54 2 Click on Next The System status dialogue box is displayed with the tab General...
Страница 132: ...7 Maintaining the product 132 2 Click on Continue The System status dialogue with the General tab is shown...
Страница 177: ...8 Tips and assistance 177 Ende der Liste f r Textmarke Inhalt...
Страница 178: ...0970 4020 en 13 V04 5 SP4 en GB...