7 Maintaining the product
140
7. Click on
[Next >]
if the probe should also be included in the
configuration or
8.
Select
Replace Existing Probe
if the probe should replace
another probe.
9. Click on
[Next]
and follow the instructions of the startup wizard.
Note Further configuration is carried out in the same way
as
Add New Component
, see
Adding new components,
page
134
.
Pos: 171 /TD/Produkt instand halten/testo Saveris/Kalibrierung und Justage @ 0\mod_1188476360640_79.docx @ 2882 @ 25555 @ 1
7.3.
Calibration and adjustment
Calibration
The comparison of a measured value with the correct value
under specified conditions. This includes documentation of
the deviation, correction of the measurement uncertainty
and creation of the certificate. Calibration does not include
adjustment of the measuring instrument.
Adjustment
A measuring instrument is used to measure a correct,
known value (confirmed by a superordinate body). The
instrument is then adjusted to this value.
Содержание 0572 0158
Страница 1: ...Measurement data monitoring with testo Saveris Small Business Edition Instruction manual...
Страница 2: ...2...
Страница 54: ...5 First steps 54 2 Click on Next The System status dialogue box is displayed with the tab General...
Страница 132: ...7 Maintaining the product 132 2 Click on Continue The System status dialogue with the General tab is shown...
Страница 177: ...8 Tips and assistance 177 Ende der Liste f r Textmarke Inhalt...
Страница 178: ...0970 4020 en 13 V04 5 SP4 en GB...