7 Maintaining the product
142
4. Once the probe comes back following calibration, link it back up
with the base or the converter.
5. The probe automatically acquires its former configuration data
and continues measuring.
Procedure 2
Here the probe removed from the system is temporarily replaced
with another probe in order to retain measurement data even
during the calibration and adjustment.
1. Register the replacement probe on the base.
2. Configure the replacement probe using the startup wizard. Here
choose the option
Use as a replacement
.
3. Take replacement probe to the measuring point and wait for
acclimatisation.
4. Press the connect key on both probes, one after the other.
- Replacement is now complete (measurement data is supplied
to the system via the replacement probe).
5. Carry out calibration and adjustment.
6. Register the probe on the base again.
7. Configure the probe using the startup wizard. Here, at
Connect
select the option
Off
and subsequently select the option
Use as
a replacement
.
8. Take probe back to the measuring point and wait for
acclimatisation.
9. Press the connect key on both probes, one after the other.
- Replacement is now complete (measurement data is supplied
to the system via the original probe once again).
Pos: 172 /TD/Produkt instand halten/testo Saveris/Base sichern @ 0\mod_1189416604765_79.docx @ 4183 @ 2 @ 1
7.4.
Saving data in the Saveris base
✓
Saveris base is started up and connected to the PC.
1.
Launch testo startup wizard.
2.
Enter IP address of the Saveris base.
3.
Click on
[Base backup]
on the
Projects
tab.
- Message
For backup, the base must be restarted
is
displayed.
Do not
confirm the message.
4.
Shut down the Saveris base: in the
Info Base
menu, press
[Esc]
briefly twice.
5.
Start up the Saveris base: press and hold down
[Esc]
.
- Menu
Select language
appears.
Содержание 0572 0158
Страница 1: ...Measurement data monitoring with testo Saveris Small Business Edition Instruction manual...
Страница 2: ...2...
Страница 54: ...5 First steps 54 2 Click on Next The System status dialogue box is displayed with the tab General...
Страница 132: ...7 Maintaining the product 132 2 Click on Continue The System status dialogue with the General tab is shown...
Страница 177: ...8 Tips and assistance 177 Ende der Liste f r Textmarke Inhalt...
Страница 178: ...0970 4020 en 13 V04 5 SP4 en GB...