4 Transmitter
42
Message Display
Cause
Remedying of fault
01505 Watchdog
error
Due to a processor
error, the transmitter
performs an automatic
restart.
If the problem occurs
frequently, contact
Testo Service.
01115
Low adjustment
temperature
The ambient
temperature is too low
during the pressure
adjustment.
Take necessary
measures to raise
ambient temperature,
e.g. through heating.
01116
High adjustment
temperature
The ambient
temperature is too high
during the pressure
adjustment.
Take necessary
measures to lower
ambient temperature,
e.g. through venting.
Pos: 66 /TD/Produkt verwenden/MUF 63xx/Status-Warnmeldungen/Behandlung von Alarmmeldungen 635x @ 3\mod_1236081852932_79.docx @ 27252 @ 3 @ 1
4.5.4.
Handling alarm messages
Shown on the
display
3
Can be used for
collective alarm
4
Additional
message end
5
New limit value
x
Scaling changed
x
Adjustment DeltaP
x
Pressure too high
x
x
Alarm 1
x
Alarm 2
x
Alarm 3
x
Alarm 4
x
Transmitter reset
x
3
If multiple messages/alarms are activated at the same time, only the last
message/alarm is shown. If this is cancelled, the other messages that are still
active are no longer displayed.
4
The message can be assigned the collective alarm function, which means that
the collective alarm is activated as soon as at least one of the messages
assigned to it is activated. The collective alarm can be assigned to each of the 4
optional relays. The collective alarm is then always the same, as it can only be
defined once.
5
The message is shown upon the occurrence of the event causing the message
as well as when closing. Two entries appear in the history in the P2A software:
Message text_start and Message text_end.
Содержание 0555 6383
Страница 2: ...2 ...
Страница 29: ...4 Transmitter 29 ...
Страница 74: ...6 Tips and assistance 74 ...
Страница 75: ...6 Tips and assistance 75 ...
Страница 76: ...0970 6384 en 02 V01 12 V01 70 1 en ...