4 Transmitter
25
Pos: 47 /TD/Erste Schritte/MUF 63xx/n-Punkt-Abgleich _635x @ 3\mod_1236352382511_79.docx @ 27522 @ 4 @ 1
4.3.3.2.
n-point adjustment
With an n-point adjustment, the parameters at the 3-6
measurement points are adjusted to the reference value. The
reference conditions are obtained by using a precise pressure
sensor that should be 5-times more accurate than the transmitter.
1
Positive pressure connection
2.
Negative pressure connection
3.
Pressure sensor
The number of measuring points is set to 3 by the factory
and can only be changed using the P2A software (see
n-
point adjustment
, page 63).
The n-point adjustment must always be carried out to its full
extent and in good time at all selected measurement points.
✓
A precise pressure sensor (5-times more accurate than the
transmitter, e.g. DPC precision pressure sensor from testo
industrial services) is available.
1.
Connect the positive output of the pressure sensor (
3
) to the
positive pressure connection of the transmitter (
1
) and the
negative output of the pressure sensor (
3
) to the negative
pressure connection of the transmitter (
2
).
2.
Transfer the reference pressure value created with the pressure
sensor into the P2A software (see n-point adjustment, page 63)
or enter it via the user menu (see Performing analog
adjustment, page 37).
3.
Repeat step 2 for all of the measuring points.
4.
Disconnect connections between the pressure sensor and the
pressure connections of the testo 6383.