4 Transmitter
35
3.
Accept setting by pressing
SET
and test with multimeter
(minimum requirement: resolution 6.5 digits, at least 2-times
more accurate than the 6383) as follows:
Analog output 1: Via test contacts under service flap, see
diagram.
1
Channel 1
test contacts
2
Service
interface
3
Multimeter
4.
Return to
Test Analog Output
with
ESC
and use
or
to
continue to
Test Relay Output
.
Testing functionality of the pressure sensor (
Test pressure
sensor
)
This function is only required to calibrate the pressure
sensor.
Testing functionality of relay outputs
1.
Press
SET
and choose between
Alarm 1, 2, 3, 4
with
or
.
2.
Press
SET
.
The relay can now be tested. You can choose between
OFF
and
ON
using
or
. If
ON
is chosen, the NO contact is
closed, the NC contact opened. If
OFF
is chosen, the NC
contact is closed, the NO contact opened.
3.
To test, route a measuring cable from the relay terminals (see
Connecting the relay outputs, page
19
) out of the transmitter to
a multimeter (resistance measurement) or continuity tester.
4.
Return to
Test Relay Output
with
SET
(starts relay test) or
ESC
(exits the menu without relay test).
Reading off min./max. values of channels
To reset the max./min. values, see Editing Reset main menu, page
39
1.
Read off the min./max. values of the channel in succession with
or
and return to the
Main Menu Analysis
with
ESC
.
Содержание 0555 6383