4 Transmitter
43
Shown on the
display
3
Can be used for
collective alarm
4
Additional
message end
5
Analog adjustment
x
T ambient high
x
x
T ambient low
x
x
Supply voltage low
x
x
Watchdog error
x
Perform the
Confirm message
function (acknowledgement of the
alarm via the control keys on the transmitter):
• The message/alarm is no longer shown on the display and the
optical alarm goes out, where applicable. If multiple
messages/alarms are active at the same time, all are reset
simultaneously.
• If at least one message is assigned to the collective alarm, the
collective alarm is reset. If the collective alarm is set on a relay,
the relay is also reset, meaning switched to its neutral position.
Pos: 67 /TD/Produkt verwenden/MUF 63xx/Status-Warnmeldungen/Namur Fehlerbedingungen 635x @ 3\mod_1236939722916_79.docx @ 29433 @ 3 @ 1
4.5.5.
Namur fault conditions
If the faults named in the following table occur, the analog outputs
output special values that enable a general fault warning in the
higher-level control system. The values correspond to the "Namur"
industry standard.
Status
message in
the display
Class
Display
value in the
display
Analog output
0 to
20 mA
4 to
20 mA
1 V
5 V 10 V
Watchdog
error
Error
Previous
value stops
21 mA
3.8 mA 1.1 V 5.5 V 11 V
Value below
min. scale
Underrange Reading
0 mA
3.8 mA 0 V
0 V 0 V
Value above
max. scale
Overrange Reading
20.5 mA 20.5 mA 1.1 V 5.5 V 11 V
Pressure too
high
Overrange ooooo ·
20.5 mA 20.5 mA 1.1 V 5.5 V 11 V