Operation
Vision G2 Testers User Guide—74-108-800 Rev. H—12 June 2019
84 of 158
Rinse Volume: The amount of volume pumped before detection
or collection is done. The range is 0-99.9 mL. This volume
should be large enough to ensure that old material in the lines
is rinsed past the needles of the collector and/or UV cells to
ensure the sample collected or detected is representative of
what is in the vessels of the tester. The default value is 7 mL.
NOTE: If a Vision AutoFilter is installed, this volume is pumped
without a filter in place.
Rinse Volume with filter: This field is only visible if a Vision
AutoFilter is installed. This is the sample rinse volume pumped
with a filter in place. The range is 0-99.9 mL. A minimum of 3
mL is recommended. This option is not visible if the AutoFilter
option in the protocol is set to No. A filter icon is present to
indicate this volume is pumped with a filter in place.
Collect Volume: The volume collected into test tubes or vials
during a sampling cycle. The range is 0-99.9 mL. The maximum
for 16 x 100 test tubes is recommended to be no more than 15
mL; for 12 x 32 vials, no more than 1.5 mL. This function only
applies to Collect Only and Collect and Detect modes.
Media Replace: Volume from the media replace source
(typically Source C if configured) to be transferred to the current
source after sampling. The range is 0-99.9 mL.
If the AutoFilter is configured for pull-through, an option is
available to select whether or not the AutoFilter will keep a filter
in place for the Media Replace step. If enabled an icon of a filter
will appear in the Media Replace field.
Generally, the Media Replace volume should be set to be the
same as the total of the Rinse Volume and the Collect Volume.
For systems that use a Retrieval Reservoir, the media replace
volume should be same as the Collect Volume.
Detect Volume: The volume pumped after detection by a UV
spectrophotometer. Only applicable with Collect and Detect
mode. The range is 0-99.9 mL. This volume must be less than
the Collect Volume.
Read Delay: This is the time required to hold the flow cell full for
detection, and should be programmed to the time required to
analyze all the flow cells used. It is the length of the pause that
occurs between the collection volume and detect volume. This
option is only applicable with Collect and Detect mode. The
range is 0 to 12,000 seconds.
Содержание VISION G2 CLASSIC 6
Страница 1: ...VISION G2 CLASSIC 6 AND ELITE 8 DISSOLUTION TESTERS USER GUIDE 74 108 800 Rev H 12 June 2019...
Страница 147: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 146 of 158 Appendix D Reports...
Страница 148: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 147 of 158...
Страница 149: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 148 of 158...
Страница 150: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 149 of 158...
Страница 151: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 150 of 158...
Страница 152: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 151 of 158...
Страница 153: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 152 of 158...
Страница 154: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 153 of 158...