Operation
Vision G2 Testers User Guide—74-108-800 Rev. H—12 June 2019
65 of 158
The Alarms button at the bottom also brings up the Edit Alarms
screen. There is a tab available for each alarm; the function of
each alarm is listed below.
The Reset button at the bottom of the screen will clear or turn
off the selected alarm.
Preheat Alarm screen
This screen allows the user to enable a preheat wakeup alarm.
The waterbath heats to the set point at the specified time and
date. This allows the tester to be ready for the start of a workday
while saving power overnight.
Bath Temperature: This is the temperature that the tester will
heat up to. Brings up numeric keypad. Range 25-55 °C; preset
values are 32, 37, and 55 °C.
Alarm Time: The time the alarm will begin the bath heating.
Alarm Date: The date that the alarm will begin the tester heating
(or Disabled).
Содержание VISION G2 CLASSIC 6
Страница 1: ...VISION G2 CLASSIC 6 AND ELITE 8 DISSOLUTION TESTERS USER GUIDE 74 108 800 Rev H 12 June 2019...
Страница 147: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 146 of 158 Appendix D Reports...
Страница 148: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 147 of 158...
Страница 149: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 148 of 158...
Страница 150: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 149 of 158...
Страница 151: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 150 of 158...
Страница 152: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 151 of 158...
Страница 153: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 152 of 158...
Страница 154: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 153 of 158...