Operation
Vision G2 Testers User Guide—74-108-800 Rev. H—12 June 2019
92 of 158
(S) Dispense Flow Rate: The parameter is speed of the syringes
while dispensing. This changes the speed at which the Vision
G2 AutoPlus syringes dispense material. The range is 0.6 to
23.8 mL/min. The default setting is 19.0 mL/min.
NOTE: 19.0 mL/min. should not be exceeded, or the Vision G2
AutoPlus could be damaged.
(T) Drive Speed: The parameter is the tester drive speed. It
allows the user to set a new motor speed for the dissolution
tester. The range is 0 for off, or 25 to 250 rpm.
(F) Filtered Media Replace: Toggles whether or not the media
replace is done with a filter in place. The options are yes and
no.
(F) Filtered Retrieval: Toggles whether or not the retrieval
volume is pumped with a filter in place. The options are yes and
no.
(S) Load Flow Rate: The parameter is speed of the syringes
while loading. This changes the speed at which the Vision G2
AutoPlus syringes load material. The range is 0.6 to 23.8
mL/min. The default setting is 19.0 mL/min.
NOTE: 19.0 mL/min. should not be exceeded, or the Vision G2
AutoPlus could be damaged.
(B) Mag Home Offset: (Version 2.64 and above only) Allows the
mag home offset position to be set as a pre-run event. The
offset remains active for the duration of the test but user
interaction such as printing the test report or acknowledging the
test over prompt clears the offset position. The range is 0-30
mm.
(T) Magazine Travel: This parameter defines magazine stroke
from home to the sampling position. The range is 0 (disabled)
to 100 mm (for 1 L vessels) or 140 mm (for 2 L vessels). The
default value is 60 mm. Preset options for magazine travel are
Baskets 500, Baskets 900, Paddles 500, Paddles 900, Small
Vol, and Chinese Small Vol for selecting predetermined
sampling height based on apparatus and commonly used media
volume in the vessels.
For example, Baskets 500 would be a preset value for the
proper magazine travel in a test involving 500 mL of media in
each vessel.
(S) Media Replace: The parameter is volume, and the range is
0 to 99.9 mL. This changes the media replace volume pumped
during the test.
Содержание VISION G2 CLASSIC 6
Страница 1: ...VISION G2 CLASSIC 6 AND ELITE 8 DISSOLUTION TESTERS USER GUIDE 74 108 800 Rev H 12 June 2019...
Страница 147: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 146 of 158 Appendix D Reports...
Страница 148: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 147 of 158...
Страница 149: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 148 of 158...
Страница 150: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 149 of 158...
Страница 151: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 150 of 158...
Страница 152: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 151 of 158...
Страница 153: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 152 of 158...
Страница 154: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 153 of 158...