Operation
Vision G2 Testers User Guide—74-108-800 Rev. H—12 June 2019
73 of 158
Adjust Temp Probe Offset screen
Touch the Cal Temp button at the bottom of the screen for the
Adjust Temp Probe Offset screen. This screen allows the user
to adjust the offsets to each of the temperature probes installed
on the tester and within the Vision heater.
There are two tabs, Instrument Probes and Vessel Probes. To
adjust the offset of a probe, touch the field in which it is listed
and a numeric keypad will appear. Enter the correct
temperature as measured by an NIST traceable thermometer to
make the correction.
NOTE: If “--.--“ is displayed in any of the fields, that indicates
there is an issue with that temperature probe. Contact Technical
Support for assistance.
Instrument Probes tab
External: The external temperature probe installed on the
dissolution tester.
Heater: The temperature probe located in the heater which
controls the rate at which the heater is heating the bath water.
Содержание VISION G2 CLASSIC 6
Страница 1: ...VISION G2 CLASSIC 6 AND ELITE 8 DISSOLUTION TESTERS USER GUIDE 74 108 800 Rev H 12 June 2019...
Страница 147: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 146 of 158 Appendix D Reports...
Страница 148: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 147 of 158...
Страница 149: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 148 of 158...
Страница 150: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 149 of 158...
Страница 151: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 150 of 158...
Страница 152: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 151 of 158...
Страница 153: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 152 of 158...
Страница 154: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 153 of 158...