Troubleshooting
Vision G2 Testers User Guide—74-108-800 Rev. H—12 June 2019
126 of 158
Mechanical Troubleshooting
Problem
Possible Causes
Recommended Action
Excessive vibration in
system
Belts are too tight or too
loose.
Replace or adjust belts
according to maintenance
procedure in this manual.
Tester is not level.
Ensure all feet of tester are
secure on bench.
Alignment nut of high stand
tester may be loose.
Tighten alignment nut.
Paddle/Basket heights not
correct
Improper operation
Confirm that spindle clamp
and spindle are properly
engaged.
Confirm that paddle or basket
shaft is properly tightened in
spindle shaft.
Clamp has loosened or
spindle shaft replaced.
Perform the spindle height
adjustment procedure listed in
the maintenance section.
Содержание VISION G2 CLASSIC 6
Страница 1: ...VISION G2 CLASSIC 6 AND ELITE 8 DISSOLUTION TESTERS USER GUIDE 74 108 800 Rev H 12 June 2019...
Страница 147: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 146 of 158 Appendix D Reports...
Страница 148: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 147 of 158...
Страница 149: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 148 of 158...
Страница 150: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 149 of 158...
Страница 151: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 150 of 158...
Страница 152: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 151 of 158...
Страница 153: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 152 of 158...
Страница 154: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 153 of 158...