Maintenance
Vision G2 Testers User Guide—74-108-800 Rev. H—12 June 2019
122 of 158
7.
Slide the tensioner to add tension to the drive belt.
Proper tension will be when the belt can be gently
deflected 6-10 mm between spindles 5 and 6 on the
Elite 8, or spindles 2 and 3 on the Classic 6.
8.
Replace the cover by aligning it over the spindles
and monument and then pressing down until it pops
back into place.
9.
Reinstall the spindle shafts.
Heater and Temperature Probe Adjustment
1.
Place an NIST traceable thermometer next to the
temperature probe.
2.
From the home screen, touch the Menu button
followed by the Tools button.
3.
Touch the Cal Temp button at the bottom of the
screen.
4.
Locate the probe on the Instrument or Vessel
Probes tab and touch that field.
5.
Use the numeric keypad to enter the temperature on
the NIST traceable thermometer, then touch OK.
Содержание VISION G2 CLASSIC 6
Страница 1: ...VISION G2 CLASSIC 6 AND ELITE 8 DISSOLUTION TESTERS USER GUIDE 74 108 800 Rev H 12 June 2019...
Страница 147: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 146 of 158 Appendix D Reports...
Страница 148: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 147 of 158...
Страница 149: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 148 of 158...
Страница 150: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 149 of 158...
Страница 151: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 150 of 158...
Страница 152: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 151 of 158...
Страница 153: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 152 of 158...
Страница 154: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 153 of 158...