Appendix A: Errors
Vision G2 Testers User Guide—74-108-800 Rev. H—12 June 2019
140 of 158
Vision Heater Errors
Error
Code
Error
Category Title
Message
Fault
Check the Following
1500
4
Heater Comm. Error
The Vision Heater is not
communicating. Make sure the Vision
Heater is online and connected.
The tester and heater are not
communicating.
1. Vision Heater power
2. Vision Heater connection
1501
6
Heater Control Test Error
The tester external temperature
probe did not record a rise in
temperature during a heating test.
1. External probe in bath water
2. Bath is full
3. Heater is heating
1502
4
Heater Probe Error
Vision Heater cannot communicate
with Inlet temperature probe. Contact
Tech Support.
Vision Heater cannot communicate
with Inlet temperature probe.
1. Inlet probe connection
2. Inlet probe failure
3. Heater board failure
1503
4
Heater Power Error
Vision Heater coil is not receiving
power. Turn off the heater for 10 min. If
the problem persists, contact Tech
Support.
Heater coil is not receiving power.
1. Emergency thermo-switch has tripped
breaking the connection between the heater
and power source.
2. Heater coil connection
3. Heater coil failure
4. Heater board failure
1504
4
Safety Probe Error
Vision Heater cannot communicate
with safety probe. Contact Tech
Support.
Heater cannot communicate with
safety probe.
1. Safety probe connection
2. Safety probe failure
3. Heater board failure
1505
6
Heater Comm. Shutdown
1506
4
Heater Temp Rise Error
Vision Heater is heating up too quickly.
Make sure the Vision Heater is primed.
Heater is heating up too quickly.
The safety probe has been
triggered.
1. Priming of the pump
2. Safety probe accuracy
3. Heater board failure
1507
6
Heater Probe Bad Reading
1508
6
Safety Probe Bad Reading
1509
4
Heater High-Temp Error
Vision Heater High-Temp Error
1510
4
Heater High-Current Error
Vision Heater High-Current Error
1511
4
Heater Maintenance Log ErroVision Heater Maintenance Log Error
1512
4
Heater Probe Memory Error
Vision Heater Inlet temperature probe
memory corruption. Contact Tech
Support.
1. Inlet probe memory bad
2. Inlet probe failure
3. Heater board failure
1513
4
Safety Probe Memory Error
Vision Heater safety probe memory
corruption. Contact Tech Support.
1. Safety probe memory bad
2. Safety probe failure
3. Heater board failure
Содержание VISION G2 CLASSIC 6
Страница 1: ...VISION G2 CLASSIC 6 AND ELITE 8 DISSOLUTION TESTERS USER GUIDE 74 108 800 Rev H 12 June 2019...
Страница 147: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 146 of 158 Appendix D Reports...
Страница 148: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 147 of 158...
Страница 149: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 148 of 158...
Страница 150: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 149 of 158...
Страница 151: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 150 of 158...
Страница 152: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 151 of 158...
Страница 153: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 152 of 158...
Страница 154: ...Appendix D Reports Vision G2 Testers User Guide 74 108 800 Rev H 12 June 2019 153 of 158...