Dissolution Testing
Vision G2 Testers User Guide—74-108-800 Rev. H—12 June 2019
114 of 158
c. Paddles
i. Remove the vessel cover over each of
the positions.
ii. Rapidly drop a dosage in each
position.
iii. Touch Run to start the test.
iv. Replace the vessel covers over each
position.
d. Paddles with ADD cover
i. Place the dosage in each of the dose
tubes located on the ADD cover for all
positions.
ii. Place the dose caps on top of the dose
tubes for all positions.
iii. Drop the dosage into all positions by
depressing the push pins on each
cover.
iv. Touch Run to start the test.