ACCQPrep HP125
Section 5 PeakTrak
5-24
• Offest - Offset is the voltage applied to all masses. Large
values tend to increase fragmentation but reduce adduct
formation.
• Span - Span voltage defines an increased voltage
applied as the mass increases. As with the offset, larger
values increase fragmentation.
• Close - Closes the window.
5.2.14 Manual Control
Manual Control can assist with method development, mainte-
nance of the system plumbing, and system troubleshooting. The
Manual Control window can be opened by selecting T
OOLS
>
M
ANUAL
C
ONTROL
from the PeakTrak menu. The Manual
Control Window has several controls:
• Solvent Selections – If the Solvent Selector Valve
Module is installed, you can select which solvent is
pumped during manual pump operation.
• Percentage Solvent B – type or select the mixture
percentage.
• Flow Rate – Type or select the pump’s flow rate in
mL/min.
• Pump into Tube # – As a default setting, solvent is
pumped to the waste port during manual control. If you
would like to collect the solvent in a tube, select the
O
PTION
button and enter the tube number. This feature
can be useful when trying to recover a compound that
has precipitated or “crashed” somewhere in the fluid
path.
• Prime A – Click the “P
RIME
A” button to pump 100% of
solvent A at the selected flow rate.
• Prime B – Click the “P
RIME
B” button to pump 100% of
solvent B at the selected flow rate.
• P
RIME
XX
% B – Click this button to pump a mix of
Solvent A and B.
• S
TOP
– Click this button to stop the pump. This button is
only active while the pump is running.
The raw lamp energy display can be used to verify flow cell
cleanliness. If the energy is low, it can be monitored while
pumping a solvent to clean the flow cell. Click the C
LOSE
M
ANUAL
C
ONTROL
button to close the window.
5.2.15 Configuration
Network Configuration –
•Network configurations settings
require assistance from your network administrator.
Contact your Information Technology department before
changing the settings. For more information on Network
Configuration, see Technical Note 28 available at
www.isco.com.
Instrument IP Address Configuration –
•Use these
settings to specify the instrument’s network address and
its parameters:
Network Type –
•Select the “Static IP” option.
Содержание ACCQPrep HP125
Страница 2: ......
Страница 6: ...ACCQPrep HP125 Installation and Operation Guide Safety Warnings 6 ...
Страница 10: ...ACCQPrep HP125 Table of Contents 4 ...
Страница 16: ...ACCQPrep HP125 Section 1 Introduction 1 6 ...
Страница 100: ...ACCQPrep HP125 Section 5 PeakTrak 5 36 ...
Страница 128: ...ACCQPrep HP125 Section 6 Maintenance 6 28 ...
Страница 130: ...ACCQPrep HP125 Appendix A A 2 This page is intentionally left blank ...
Страница 132: ...ACCQPrep HP125 Appendix A A 4 This page is intentionally left blank ...
Страница 134: ...ACCQPrep HP125 Appendix A A 6 This page is intentionally left blank ...
Страница 135: ...A 7 Appendix A ACCQPrep HP125 This page is intentionally left blank ...
Страница 136: ...A 8 This page is intentionally left blank ...
Страница 138: ......
Страница 140: ......
Страница 142: ......