ACCQPrep HP125
Section 6 Maintenance
6-18
6.6 AutoSampler
Maintenance
The Peristaltic pump on the back of the AutoSampler may need
to be replaced.
Peristaltic Tubing
Replacement
The peristaltic tubing P/N SP7477
.
Figure 6-25 Back of the AutoSampler
1. Removed the outer band of the peristaltic pump (Figure
6-26).
Figure 6-26 Removing the outer band
2. Lift up to remove the peristaltic tubing from the pump,
repeat on the bottom (Figure 6-27).
Figure 6-27 Removing the tubing