1.5 Specification
Table 1-1 ACCQ
Prep
HP125
Dimensions of ACCQ
Prep
(Footprint without column or column cover
installed)
27.5 x 14.0 x 20.0 in (69.9 x 35.6 x 50.8 cm)
Dimensions of ACCQ
Prep
AS 2x1
18.9 x 14.0 x 19.3 in (48.0 x 35.6 x 49.0 cm)
Weight
(Including AutoInjector, Column Selector
Valve Module, Solvent Selector Valve Mod-
ule, and column cover)
93.2 lbs (42.3 kg)
106 lbs (48 kg) with optional Evaporative Light Scattering Detector
(ELSD)
Weight of ACCQ
Prep
AS 2x1
27.3 lbs (12.4 kg)
User Interface
15" touchscreen
Power
Input voltage range from 100 to 240 VAC,
300 VA maximum.
Line cord is the disconnect device. Power connection is via IEC
60320 C14 power inlet.
Line Frequency
50/60 Hz
Ambient Temperature
20 to 40 °C (maximum temperature must be at least 15 °C below the
boiling point of the solvent)
Humidity (when connected to power)
90% relative humidity maximum at 20 to 40 °C (non-condensing)
Flow Rate Range
1-125 mL/min, ELSD peak collection limited to 4-125 mL/min
Flow Rate Accuracy (tested with water at 6.9
bar or 100 psi)
± 2%
Pressure Limit
414 bar (6000 psi) derated linearly from 6000 psi at 75 mL/min to
4000 psi at 125 mL/min
Pressure Accuracy
5% of full scale
Gradient Formation
Binary gradient 2 solvent inlets. Optional Solvent Selector Valve Mod-
ule expands to 3 solvents each for A & B
Gradient Accuracy:
±1% of full scale
Peak Detection Modes
Slope or threshold
Flow Cell Pathlength
0.3
mm, ±25% (Standard, other pathlengths available)
UV Detection Wavelength
200 to 400 nm, optional 200 to 800 nm UV-Vis
Wavelength Accuracy
±5 nm
Fraction Accuracy
±[2mL
+
(flow rate
÷
60)]
Optional ELSD
Gas Inlet Pressure
60 to 70 psig
Gas Consumption
<2.5 SLPM
Spray Chamber Temperature
Setting range: 10 to 60 °C
limited to minimum of 5 °C below ambient
Drift Tube Temperature
Setting range: 30 to 90 °C
Must be 5 °C above spray chamber temperature Minimum tempera-
ture is 5 °C above ambient Maximum temperature is 60 °C above
ambient
Содержание ACCQPrep HP125
Страница 2: ......
Страница 6: ...ACCQPrep HP125 Installation and Operation Guide Safety Warnings 6 ...
Страница 10: ...ACCQPrep HP125 Table of Contents 4 ...
Страница 16: ...ACCQPrep HP125 Section 1 Introduction 1 6 ...
Страница 100: ...ACCQPrep HP125 Section 5 PeakTrak 5 36 ...
Страница 128: ...ACCQPrep HP125 Section 6 Maintenance 6 28 ...
Страница 130: ...ACCQPrep HP125 Appendix A A 2 This page is intentionally left blank ...
Страница 132: ...ACCQPrep HP125 Appendix A A 4 This page is intentionally left blank ...
Страница 134: ...ACCQPrep HP125 Appendix A A 6 This page is intentionally left blank ...
Страница 135: ...A 7 Appendix A ACCQPrep HP125 This page is intentionally left blank ...
Страница 136: ...A 8 This page is intentionally left blank ...
Страница 138: ......
Страница 140: ......
Страница 142: ......