ACCQPrep HP125
Section 1 Introduction
1-5
Electrical Safety per EN 61010-1
Pollution Degree
2
Installation Category
II
Maximum Altitude
2000 meters
Note 1. All specifications are subject to change.
Table 1-1 ACCQ
Prep
HP125
Table 1-2 ACCQ
Prep
HP125 with PurIon System
Dimensions (H x W x D)
ACCQ
Prep
: 27.5 x 14.0 x 20.0 in (69.9 x 35.6 x 50.8 cm)
Mass Spectrometer 26 x 11 x 22 in (66 x 28 x 56 cm)
Roughing Pump 10 x 9 x 18 in (26 x 23 x 46 cm)
ELSD Detection
Option that can be combined with either UV or UV-Vis
Mass Spectrometry Detection
10 – 1200 Dalton for S model, 10-2000 for L model,
1 Dalton Resolution
Electrospray Ionization (ESI) or optional Atmospheric Pressure
Chemical Ionization (APCI)
Simultaneous positive and negative ionization for both S and L
models.
Note 1. All specifications are subject to change.
Table 1-3 Component Materials List
Chromatographic Tubing
316 stainless steel tubing, PEEK, and
Fluoropolymer
Drain Tubing
Vinyl with FEP liner
Chromatographic Valves
PEEK, PPS, perfluorelastomer (FFKM)
Flowcell
303 SST, Type ES Quartz, SIMRIZ
®
SZ485
®
Chromatography Pump
316 Stainless Steel, UHMWPE, ETFE, ruby, sapphire, and zirconia
Pressure Transducer
316 and 17-4 PH stainless steels and perfluorelastomer (FFKM)
AutoSampler Wash
Pharm-A-Line
,
polypropylene
Содержание ACCQPrep HP125
Страница 2: ......
Страница 6: ...ACCQPrep HP125 Installation and Operation Guide Safety Warnings 6 ...
Страница 10: ...ACCQPrep HP125 Table of Contents 4 ...
Страница 16: ...ACCQPrep HP125 Section 1 Introduction 1 6 ...
Страница 100: ...ACCQPrep HP125 Section 5 PeakTrak 5 36 ...
Страница 128: ...ACCQPrep HP125 Section 6 Maintenance 6 28 ...
Страница 130: ...ACCQPrep HP125 Appendix A A 2 This page is intentionally left blank ...
Страница 132: ...ACCQPrep HP125 Appendix A A 4 This page is intentionally left blank ...
Страница 134: ...ACCQPrep HP125 Appendix A A 6 This page is intentionally left blank ...
Страница 135: ...A 7 Appendix A ACCQPrep HP125 This page is intentionally left blank ...
Страница 136: ...A 8 This page is intentionally left blank ...
Страница 138: ......
Страница 140: ......
Страница 142: ......