Operating Instructions
Model 300E Carbon Monoxide Analyzer
Instruction Manual
80
P/N 04288 REV: A
6.6.2.
Analog Output Step Test
This test can be used to assure the analog outputs are calibrated and working
properly. The test forces the
A1
,
A2
and
A3
analog output channels to produce
signals ranging 0% to 100% of whatever the range in which they are set in 20%
increments. This test is useful for verifying the operation of the
datalogging/recording devices attached to the analyzer.
To begin the Analog Output Step Test press:
SAMPLE
RANGE = 50.0 PPM
CO =
< TST TST > CAL
SETUP
SETUP C.3
CFG DAS
RNGE
PASS CLK
MORE
EXIT
SETUP C.3
COMM VARS
DIAG
HALT
EXIT
Exit at Any
Time to Return
to main the
Sample Display
DIAG
SIGNAL I / O
NEXT
ENTR
EXIT
Exit 2X’s
to Return
to the
Diag Menu
SETUP C.3
ENTER DIAG PASS: 818
8 1 8
ENTR
EXIT
DIAG
ANALOG OUTPUT
PREV
NEXT
ENTR
EXIT
DIAG AOUT
ANALOG
OUTPUT
0%
EXIT
Performs
Analog Output
step test.
0% - 100%
DIAG AOUT
ANALOG OUTPUT
[0%]
EXIT
Pressing the key under the STEP % will
halt the Test.
Brackets will appear Around the Step %:
EXAMPLE [20]
Pressing the same key again will resume
the Test.
Содержание 300E
Страница 40: ...Getting Started Model 300E Carbon Monoxide Analyzer Instruction Manual 40 P N 04288 REV A INTENTIONALLY BLANK ...
Страница 122: ...Operating Instructions Model 300E Carbon Monoxide Analyzer Instruction Manual 122 P N 04288 REV A INTENTIONALLY BLANK ...
Страница 186: ...THEORY OF OPERATION Model 300E Carbon Monoxide Analyzer Instruction Manual 186 P N 04288 Rev A INTENTIONALLY BLANK ...
Страница 224: ...Model 300E Carbon Monoxide Analyzer Instruction Manual Appendix A P N 04288 REV A INTENTIONALLY BLANK ...
Страница 226: ...Model 300E Carbon Monoxide Analyzer Instruction Manual Appendix B P N 04288 REV A INTENTIONALLY BLANK ...
Страница 228: ...Model 300E Carbon Monoxide Analyzer Instruction Manual Appendix C P N 04288 REV A INTENTIONALLY BLANK ...
Страница 230: ...Model 300E Carbon Monoxide Analyzer Instruction Manual Appendix D P N 04288 REV A INTENTIONALLY BLANK ...
Страница 236: ......
Страница 252: ......