Model 300E Carbon Monoxide Analyzer
Instruction Manual
TROUBLESHOOTING & REPAIR PROCEDURES
P/N 04288 REV: A
213
11.5.7.3. Analog Outputs: Current Loop
To verify that the Analog outputs with the optional current mode output are
working properly, connect a 250 ohm resistor across the outputs and use a
voltmeter to measure the output as described in Section 6.5.2 and then perform an
Analog Output Step Test as described in Section 6.6.2.
For each step the output should be within 1% of the nominal value listed in the
table below.
Table 11-9: Analog Output Test Function - Nominal Values Current Outputs
Output Range
2 -20
4 -20
Nominal Output Values
Step
%
Current
V(250 Ohms)
Current
V(250 Ohms)
1 0 2
mA 0.5V 4
1
2 20 5.6
1.4
7.2 1.8
3 40 9.2
2.3
10.4 2.6
4 60 12.8 3.2
13.6 3.4
5 80 16.4 4.1
16.8 4.2
6 100 20
5
20
5
11.5.7.4. Status Outputs
The procedure below can be used to test the Status outputs:
18. Connect a jumper between the “D“ pin and the “
s
” pin on the status output
connector.
19. Connect a 1000 ohm resistor between the “+” pin and the pin for the status
output that is being tested.
20. Connect a voltmeter between the “
s
” pin and the pin of the output being
tested (see table below).
21. Under
the
DIAG:Signal I/O
menu (see Section 11.1.3), scroll through the
inputs and outputs until you get to the output in question. Alternately turn on
and off the output noting the voltage on the Voltmeter, it should vary between
0 volts for ON and 5 volts for OFF.
Содержание 300E
Страница 40: ...Getting Started Model 300E Carbon Monoxide Analyzer Instruction Manual 40 P N 04288 REV A INTENTIONALLY BLANK ...
Страница 122: ...Operating Instructions Model 300E Carbon Monoxide Analyzer Instruction Manual 122 P N 04288 REV A INTENTIONALLY BLANK ...
Страница 186: ...THEORY OF OPERATION Model 300E Carbon Monoxide Analyzer Instruction Manual 186 P N 04288 Rev A INTENTIONALLY BLANK ...
Страница 224: ...Model 300E Carbon Monoxide Analyzer Instruction Manual Appendix A P N 04288 REV A INTENTIONALLY BLANK ...
Страница 226: ...Model 300E Carbon Monoxide Analyzer Instruction Manual Appendix B P N 04288 REV A INTENTIONALLY BLANK ...
Страница 228: ...Model 300E Carbon Monoxide Analyzer Instruction Manual Appendix C P N 04288 REV A INTENTIONALLY BLANK ...
Страница 230: ...Model 300E Carbon Monoxide Analyzer Instruction Manual Appendix D P N 04288 REV A INTENTIONALLY BLANK ...
Страница 236: ......
Страница 252: ......