Operating Instructions
Model 300E Carbon Monoxide Analyzer
Instruction Manual
76
P/N 04288 REV: A
To adjust the a zero and span signal levels of the Current Outputs, press:
EXAMPLE
Exit to Return
to the main
Sample Display
DIAG
ANALOG I / O CONFIGURATION
PREV
NEXT
ENTR
EXIT
DIAG AIO
AIN A/C FREQUENCY: 60 HZ
SET>
EDIT
EXIT
DIAG AIO
AOUT CALIBRATED: NO
< SET
SET>
CAL
EXIT
FROM ANALOG I/O CONFIGURATION MENU
DIAG AIO
CONC_OUT_2 RANGE: CURR
<SET
SET>
EDIT
EXIT
DIAG AIO
CONC_OUT_2 CALIBRATED: NO
< SET
CAL
EXIT
DIAG AIO
CONC_OUT_2 ZERO: 0 mV
U100
UP10
UP DOWN DN10 D100 ENTR EXIT
DIAG AIO
CONC_OUT_CURR, NO CAL
< SET SET>
EDIT
EXIT
Press
SET>
to select the Analog Output
channel to be configured:
DISPLAYED AS = CHANNEL
CONC_OUT_1 =
A1
CONC_OUT_2 =
A2
TEST
OUTPUT =
A3
Then Press
EDIT
to continue
DIAG AIO
AIN CALIBRATED: NO
SET>
EDIT
EXIT
ENTR r
eturns
to the previous
menu
These keys
increment/decrement the
ZERO/SPAN D-to-A converter
output by
100, 10 or 1 counts
respectively.
The resulting change in output
voltage is dispalyed on the
TOP Line.
Continue adjustments until the
correct current level is
measured on a current meter
placed in series with the
output between the Model
300E and the recording
device.
DIAG AIO
CONC_OUT_2 ZERO: 27 mV
U100 UP10
UP DOWN DN10 D100
ENTR EXIT
EXIT
ignores the
new setting.
ENTR
accepts the
new setting.
DIAG AIO
CONC_OUT_2 CALIBRATED: YES
< SET
CAL
EXIT
DIAG AIO
CONC_OUT_2 SPAN: 10000 mV
U100 UP10
UP DOWN DN10
D100
ENTR EXIT
EXAMPLE
DIAG AIO
CONC_OUT_2 ZERO: 9731 mV
U100 UP10
UP DOWN DN10 D100
ENTR EXIT
Содержание 300E
Страница 40: ...Getting Started Model 300E Carbon Monoxide Analyzer Instruction Manual 40 P N 04288 REV A INTENTIONALLY BLANK ...
Страница 122: ...Operating Instructions Model 300E Carbon Monoxide Analyzer Instruction Manual 122 P N 04288 REV A INTENTIONALLY BLANK ...
Страница 186: ...THEORY OF OPERATION Model 300E Carbon Monoxide Analyzer Instruction Manual 186 P N 04288 Rev A INTENTIONALLY BLANK ...
Страница 224: ...Model 300E Carbon Monoxide Analyzer Instruction Manual Appendix A P N 04288 REV A INTENTIONALLY BLANK ...
Страница 226: ...Model 300E Carbon Monoxide Analyzer Instruction Manual Appendix B P N 04288 REV A INTENTIONALLY BLANK ...
Страница 228: ...Model 300E Carbon Monoxide Analyzer Instruction Manual Appendix C P N 04288 REV A INTENTIONALLY BLANK ...
Страница 230: ...Model 300E Carbon Monoxide Analyzer Instruction Manual Appendix D P N 04288 REV A INTENTIONALLY BLANK ...
Страница 236: ......
Страница 252: ......