Model 300E Carbon Monoxide Analyzer
Instruction Manual
Operating Instructions
P/N 04288 REV: A
71
The Test Functions available to be reported are:
Table 6–5: Test Functions available to the TEST Analog Output Channel
Test Channel
Zero
Full Scale
NONE
Test Channel is turned off
CO MEASURE
0 mV
5000 mV*
CO REFERENCE
0 mV
5000 mV*
SAMPLE PRESS
0 "Hg
40 "Hg
SAMPLE FLOW
0 cc/m
1000 cc/m
SAMPLE TEMP
0
°
C 70
°
C
BENCH TEMP
0
°
C 70
°
C
WHEEL TEMP
0
°
C 70
°
C
CHASSIS TEMP
0
°
C 70
°
C
PHOTO DET
0 mV
5000 mV
* This refers to the internal voltage level of the function NOT the output
signal level of the Test channel itself.
Once a function is selected, the instrument not only begins to output a signal on the
A3
Analog Output, but a TEST parameter is added to the list of Test Functions
viewable via the Front Panel Display. To activate the TEST channel and select a Test
Function press:
6.4.8.
Using the Dilution Option
The Dilution Ration Option is a software option that allows the user to compensate
for any dilution of the Sample Gas before it arrives at the Sample Inlet on the Rear
Panel of the instrument. Use of this feature requires that the degree of dilution be
known.
Using the Dilution Ration Option is a 4-Step process:
STEP 1: SELECT UNITS OF MEASURE – Follow the procedure defined in
Section 6.4.6.
STEP 2: SELECT THE RANGE – Use the same procedures as described in Sections
6.4.1 through 6.4.5. Make sure that the SPAN Value entered is the
maximum expected concentration of the
UN-DILUTED
calibration gas
and that the gas is input through the dilution probe.
Содержание 300E
Страница 40: ...Getting Started Model 300E Carbon Monoxide Analyzer Instruction Manual 40 P N 04288 REV A INTENTIONALLY BLANK ...
Страница 122: ...Operating Instructions Model 300E Carbon Monoxide Analyzer Instruction Manual 122 P N 04288 REV A INTENTIONALLY BLANK ...
Страница 186: ...THEORY OF OPERATION Model 300E Carbon Monoxide Analyzer Instruction Manual 186 P N 04288 Rev A INTENTIONALLY BLANK ...
Страница 224: ...Model 300E Carbon Monoxide Analyzer Instruction Manual Appendix A P N 04288 REV A INTENTIONALLY BLANK ...
Страница 226: ...Model 300E Carbon Monoxide Analyzer Instruction Manual Appendix B P N 04288 REV A INTENTIONALLY BLANK ...
Страница 228: ...Model 300E Carbon Monoxide Analyzer Instruction Manual Appendix C P N 04288 REV A INTENTIONALLY BLANK ...
Страница 230: ...Model 300E Carbon Monoxide Analyzer Instruction Manual Appendix D P N 04288 REV A INTENTIONALLY BLANK ...
Страница 236: ......
Страница 252: ......