Model 300E Carbon Monoxide Analyzer
Instruction Manual
Specifications, Agency Approvals, and
Warranty
P/N 04288 REV: A
17
2. SPECIFICATIONS, AGENCY APPROVALS,
AND WARRANTY
2.1. Specifications
Table 2–1: Model 300E Specifications
Ranges
User selectable to any full scale range from 0-1 ppm to
0-1000 ppm
Measurement Units
ppb, ppm, µg/m
3
, mg/m
3
(user selectable)
Zero Noise
≤
0.02 ppm RMS
(1)
Span Noise
< 0.5% of reading RMS over 5 ppm
(1) (3)
Lower Detectable Limit < 0.04 ppm
(1)
Zero Drift (24 hours)
< 0.1 ppm
(2)
Zero Drift (7 days)
< 0.2 ppm
(2)
Span Drift (24 hours)
< 0.5% of reading
(2)
(4)
Span Drift (7 days)
< 1% of reading
(2)
(4)
Linearity
Better than 1% full scale
(5)
Precision 0.5%
reading
(1)
(5)
Lag Time
<10 sec
(1)
Rise/Fall Time
<60 sec to 95%
(1)
Sample Flow Rate
800 scc/min. ± 10%
Temperature Range
5 - 40
°
C operating, 10 - 40
°
C EPA Equivalency
Humidity Range
0-95% RH, Non-Condensing
Temp Coefficient
< 0.05 % per
°
C
Voltage Coefficient
< 0.05 % per V
Dimensions (HxWxD)
7" x 17" x 23.5" (178 mm x 432 mm x 597 mm)
Weight
50 lb (22.7 kg)
AC Power
100V 50/60 Hz (3.25A), 115 V 60 Hz (3.0A),
220 – 240 V 50/60 Hz (2.5A)
Environmental
Conditions
Installation Category (Over voltage Category) II Pollution Degree 2
Analog Outputs
Three (3) Outputs
Analog Output Ranges
100 mV, 1 V, 5 V, 10 V, 2-20 or 4-20 mA isolated current loop.
All Ranges with 5% Under/Over Range
Analog Output
Resolution
1 part in 4096 of selected full-scale voltage
Status Outputs
8 Status outputs from opto-isolators
Control Inputs
6 Control Inputs, 2 defined, 4 spare
Serial I/O
One (1) RS-232; One (1) RS-485 (2 connecters in parallel)
Baud Rate : 300 - 115200
Certifications
USEPA: Reference Method Number EQOA-0992-087
CE: EN61010-1:90 + A1:92 + A2:95, EN61326 - Class A
(1)
As defined by the USEPA
(2)
At constant temperature and voltage
(3)
Or 0.2 ppm, whichever is greater
(4)
Or 0.1 ppm, whichever is greater
(5)
Above 10 ppm range, otherwise 0.2 ppm for lower ranges
Содержание 300E
Страница 40: ...Getting Started Model 300E Carbon Monoxide Analyzer Instruction Manual 40 P N 04288 REV A INTENTIONALLY BLANK ...
Страница 122: ...Operating Instructions Model 300E Carbon Monoxide Analyzer Instruction Manual 122 P N 04288 REV A INTENTIONALLY BLANK ...
Страница 186: ...THEORY OF OPERATION Model 300E Carbon Monoxide Analyzer Instruction Manual 186 P N 04288 Rev A INTENTIONALLY BLANK ...
Страница 224: ...Model 300E Carbon Monoxide Analyzer Instruction Manual Appendix A P N 04288 REV A INTENTIONALLY BLANK ...
Страница 226: ...Model 300E Carbon Monoxide Analyzer Instruction Manual Appendix B P N 04288 REV A INTENTIONALLY BLANK ...
Страница 228: ...Model 300E Carbon Monoxide Analyzer Instruction Manual Appendix C P N 04288 REV A INTENTIONALLY BLANK ...
Страница 230: ...Model 300E Carbon Monoxide Analyzer Instruction Manual Appendix D P N 04288 REV A INTENTIONALLY BLANK ...
Страница 236: ......
Страница 252: ......