CALIBRATION PROCEDURES
Model 300E Carbon Monoxide Analyzer
Instruction Manual
130
P/N 04288 REV: A
SAMPLE
RANGE = 50.0 PPM CO =??.??
< TST TST > CAL
CALZ
CALS
SETUP
ZERO CAL M
RANGE = 50.0 PP CO =1.23
< TST TST >
ZERO
CONC
EXIT
Press
EXIT
to Return to the
Main
SAMPLE
Display
M-P CAL
RANGE = 50.0 PPM CO =1.23
< TST TST >
ENTR
CONC
EXIT
WAIT 10
MINUTES
Or until the
reading
stabilizes and
the
SPAN
button
is displayed
This operation changes the
calculation equations and zeroes
the instrument
This operation changes
the calculation equations
and zeroes the instrument
WARNING
!
If you are performing a
Zero CAL CHECK
DO
NOT
press
ENTR
WAIT 10
MINUTES
Or until the
reading
stabilizes and
the
ZERO
button
is displayed
If either the
ZERO
or
SPAN
buttons
fail to appear see
Section 9 for
troubleshooting
tips.
Analyzer enters
ZERO CAL
Mode
See Table 5-1 for Z/S
Valve States during
this operating mode
EXIT
halts the calibration
process and returns the
unit to
SAMPLE
Mode.
If performing a CHECK,
press
EXIT
ZERO CAL M
RANGE = 50.0 PP
CO = 0.00
< TST TST > ZERO CONC
EXIT
SAMPLE
RANGE = 50.0 PPM CO =40.58
< TST TST > CAL
CALZ
CALS
SETUP
ZERO CAL M
RANGE = 50.0 PP CO =40.58
< TST TST >
SPAN
CONC
EXIT
M-P CAL
RANGE = 50.0 PPM CO =40.00
< TST TST >
ENTR
CONC
EXIT
This operation changes the
calculation equations and zeroes
the instrument
This operation changes
the calculation equations
and zeroes the instrument
WARNING
!
If you are performing a
Span CAL CHECK
DO
NOT
press
ENTR
EXIT
halts the calibration
process and returns the
unit to
SAMPLE
Mode.
If performing a CHECK,
press
EXIT
Analyzer enters
SPAN CAL
Mode
See Table 5-1 for Z/S
Valve States during
this operating mode
ZERO CAL M
RANGE = 50.0 PP CO =40.58
< TST TST >
SPAN
CONC
EXIT
Содержание 300E
Страница 40: ...Getting Started Model 300E Carbon Monoxide Analyzer Instruction Manual 40 P N 04288 REV A INTENTIONALLY BLANK ...
Страница 122: ...Operating Instructions Model 300E Carbon Monoxide Analyzer Instruction Manual 122 P N 04288 REV A INTENTIONALLY BLANK ...
Страница 186: ...THEORY OF OPERATION Model 300E Carbon Monoxide Analyzer Instruction Manual 186 P N 04288 Rev A INTENTIONALLY BLANK ...
Страница 224: ...Model 300E Carbon Monoxide Analyzer Instruction Manual Appendix A P N 04288 REV A INTENTIONALLY BLANK ...
Страница 226: ...Model 300E Carbon Monoxide Analyzer Instruction Manual Appendix B P N 04288 REV A INTENTIONALLY BLANK ...
Страница 228: ...Model 300E Carbon Monoxide Analyzer Instruction Manual Appendix C P N 04288 REV A INTENTIONALLY BLANK ...
Страница 230: ...Model 300E Carbon Monoxide Analyzer Instruction Manual Appendix D P N 04288 REV A INTENTIONALLY BLANK ...
Страница 236: ......
Страница 252: ......