TROUBLESHOOTING & REPAIR PROCEDURES
P/N 04288 REV: A
221
35. Remove the ONE (1) screw fastening the GFC Wheel/Mask assembly to the
GFC Motor Hub.
36. Remove the GFC Wheel.
GFC FILTER MASK
GFC FILTER WHEEL
Figure 11-10: Removing the GFC Wheel
Follow the previous steps in reverse order to put the GFC Wheel/Motor assembly
back together.
11.6.3.
Disk-On-Chip Replacement Procedure
Replacing the Disk-on-Chip, may be necessary in certain rare circumstances or to
load new instrument software. This will cause all of the instrument configuration
parameters and iDAS data to be lost. However a backup copy of the operating
parameters are stored in a second non-volatile memory and will be loaded into the
new the Disk-on-Chip on power-up. To change the Disk-on-Chip follow this
procedure.
1.
Turn off power to the instrument.
2.
Fold down the rear panel by loosening the thumbscrews on each side
3.
Locate the Disk-on-Chip in the rightmost socket near the right hand side of the
CPU assembly. Remove the IC by gently prying it up from the socket.
4.
Reinstall the new Disk-on-Chip, making sure the notch in the end of the chip is
facing upward.
5.
Close the rear panel and turn on power to the machine.
Содержание 300E
Страница 40: ...Getting Started Model 300E Carbon Monoxide Analyzer Instruction Manual 40 P N 04288 REV A INTENTIONALLY BLANK ...
Страница 122: ...Operating Instructions Model 300E Carbon Monoxide Analyzer Instruction Manual 122 P N 04288 REV A INTENTIONALLY BLANK ...
Страница 186: ...THEORY OF OPERATION Model 300E Carbon Monoxide Analyzer Instruction Manual 186 P N 04288 Rev A INTENTIONALLY BLANK ...
Страница 224: ...Model 300E Carbon Monoxide Analyzer Instruction Manual Appendix A P N 04288 REV A INTENTIONALLY BLANK ...
Страница 226: ...Model 300E Carbon Monoxide Analyzer Instruction Manual Appendix B P N 04288 REV A INTENTIONALLY BLANK ...
Страница 228: ...Model 300E Carbon Monoxide Analyzer Instruction Manual Appendix C P N 04288 REV A INTENTIONALLY BLANK ...
Страница 230: ...Model 300E Carbon Monoxide Analyzer Instruction Manual Appendix D P N 04288 REV A INTENTIONALLY BLANK ...
Страница 236: ......
Страница 252: ......