Operating Instructions
Model 300E Carbon Monoxide Analyzer
Instruction Manual
106
P/N 04288 REV: A
Table 6–20: iDAS Data Channel Properties
Property Description Initial
Setting
Setting Range
NAME
The data channel’s name (for reports and RS-
232 access).
“NONE”
Up to 6 letters and
digits.
TRIGGERING
EVENT
The event that triggers this data channel to
measure and store its data parameters.
See APPENDIX A for a list of available
triggering events.
ATIMER Any
single
event
chosen form the list
LIST OF
PARAMETERS
A User-configurable list of data types to be
recorded.
See APPENDIX A for a list of available
parameters.
1
Only limited by
memory space
REPORT PERIOD The amount of time between each report.
000:01:00
000:00:01 to
366:23:59
(Days:Hours:Min.)
NUMBER OF
RECORDS
The number of reports that will be stored in the
data file.
100
1 to 999,999
(limited by available
storage space)
RS-232 REPORT Indicates whether or not a report will be
printed automatically to the RS-232 channel.
OFF
OFF or ON
CHANNEL
ENABLED
Provides a convenient means to temporarily
disable a data channel.
ON
OFF or ON
CAL. HOLD OFF Disables sampling of data parameters while
instrument is in calibration mode.
OFF
OFF or ON
6.10.3.2. Data Parameters
Data Parameters are actual types of the data may be measured and stored by the
iDAS. For each of T-API’s instrument models, the list of available data parameters is
set and non-customizable (see Appendix A for a list of available Data Parameters
for this instrument).
Most data parameters have measurement units associated with them, such as mV,
PPB, cc/m, etc., although some have no units. The iDAS is not designed to permit
these units to be changed.
Each Data Parameters has user-configurable functions that define how the data is
recorded.
Содержание 300E
Страница 40: ...Getting Started Model 300E Carbon Monoxide Analyzer Instruction Manual 40 P N 04288 REV A INTENTIONALLY BLANK ...
Страница 122: ...Operating Instructions Model 300E Carbon Monoxide Analyzer Instruction Manual 122 P N 04288 REV A INTENTIONALLY BLANK ...
Страница 186: ...THEORY OF OPERATION Model 300E Carbon Monoxide Analyzer Instruction Manual 186 P N 04288 Rev A INTENTIONALLY BLANK ...
Страница 224: ...Model 300E Carbon Monoxide Analyzer Instruction Manual Appendix A P N 04288 REV A INTENTIONALLY BLANK ...
Страница 226: ...Model 300E Carbon Monoxide Analyzer Instruction Manual Appendix B P N 04288 REV A INTENTIONALLY BLANK ...
Страница 228: ...Model 300E Carbon Monoxide Analyzer Instruction Manual Appendix C P N 04288 REV A INTENTIONALLY BLANK ...
Страница 230: ...Model 300E Carbon Monoxide Analyzer Instruction Manual Appendix D P N 04288 REV A INTENTIONALLY BLANK ...
Страница 236: ......
Страница 252: ......