Model 300E Carbon Monoxide Analyzer
Instruction Manual
TROUBLESHOOTING & REPAIR PROCEDURES
P/N 04288 REV: A
203
3.
Make sure that the expected Span Gas Concentration entered into the
instrument during calibration is correct Span gas concentration too different
from expected span value. This can be viewed via the
RNG
Menu (see
Section 6.4).
4.
Check to make sure that there is no ambient air or Zero Air leaking into Span
Gas line.
11.3.4.
Inability to Zero – No ZERO Key
1.
Confirm that there is a good source of zero air. Dilute a tank of span gas with
two the same amount of Zero Air from two different sources. If the CO
Concentration of the two measurements is different, there is a problem with
one of the sources of Zero Air.
2.
Check for leaks in the pneumatic systems as described in Section 9.5.
3.
The Internal zero air scrubber may need maintenance. This device is only
present if the analyzer has had Zero/Span Valve Options 51 or 53 installed
this.
4.
Check to make sure that there is no ambient air leaking into Zero Air line.
11.4. Other Performance Problems
Dynamic problems (i.e. problems which only manifest themselves when the
analyzer is monitoring sample gas) can be the most difficult and time consuming to
isolate and resolve. The following section provides an itemized list of the most
common dynamic problems with recommended troubleshooting checks and
corrective actions.
11.4.1.
Temperature Problems
Individual control loops are used to maintain the setpoint of the Absorption Bench,
Filter Wheel, and IR Photo-Detector temperatures. If any of these temperatures
are out of range or are poorly controlled, the M300E will perform poorly.
11.4.1.1. Box or Sample Temperature
Box Temperature
The box temperature sensor is mounted to the Motherboard and cannot be
disconnected to check its resistance. Rather check the
BOX TEMP
signal using the
SIGNAL I/O
function under the
DIAG
Menu (see Section 11.1.3). This parameter
will vary with ambient temperature, but at ~30
o
C (6-7° above room temperature)
the signal should be ~1450 mV.
Содержание 300E
Страница 40: ...Getting Started Model 300E Carbon Monoxide Analyzer Instruction Manual 40 P N 04288 REV A INTENTIONALLY BLANK ...
Страница 122: ...Operating Instructions Model 300E Carbon Monoxide Analyzer Instruction Manual 122 P N 04288 REV A INTENTIONALLY BLANK ...
Страница 186: ...THEORY OF OPERATION Model 300E Carbon Monoxide Analyzer Instruction Manual 186 P N 04288 Rev A INTENTIONALLY BLANK ...
Страница 224: ...Model 300E Carbon Monoxide Analyzer Instruction Manual Appendix A P N 04288 REV A INTENTIONALLY BLANK ...
Страница 226: ...Model 300E Carbon Monoxide Analyzer Instruction Manual Appendix B P N 04288 REV A INTENTIONALLY BLANK ...
Страница 228: ...Model 300E Carbon Monoxide Analyzer Instruction Manual Appendix C P N 04288 REV A INTENTIONALLY BLANK ...
Страница 230: ...Model 300E Carbon Monoxide Analyzer Instruction Manual Appendix D P N 04288 REV A INTENTIONALLY BLANK ...
Страница 236: ......
Страница 252: ......