Model 300E Carbon Monoxide Analyzer
Instruction Manual
Operating Instructions
P/N 04288 REV: A
67
For example: The Low range can be set for a span of 0-150 ppm while the high
range is set for 0-50 ppm.
When the instrument’s range mode is set to Dual or Auto, a second set of slope and
offset parameters is used for computing the concentration for the high range. Also,
the concentration field in the upper right hand corner of the display identifies which
concentration is currently being displayed: C1 = Low (or
A1
) and C2 = High (or
A2
).
To set the ranges press following keystroke sequence.
SAMPLE
RANGE = 50.0 PPM
CO =
< TST TST > CAL
SETUP
SETUP C.3
CFG DAS
RNGE
PASS CLK MORE
EXIT
SETUP C.3
RANGE CONTROL MENU
MODE
SET UNIT
EXIT
SETUP C.3
RANGE MODE:
SNGL
SNGL
DUAL
AUTO
ENTR EXIT
EXIT
Returns
to the Main
SAMPLE
Display
SETUP C.3
RANGE MODE:
DUAL
SNGL
DUAL AUTO
ENTR
EXIT
SETUP C.3
RANGE CONTROL MENU
MODE
SET
UNIT
EXIT
SETUP C.3
HIGH RANGE:
50.0 Conc
0 0
1
5 0 .0
ENTR
EXIT
SETUP C.3
LOW RANGE:
150.0 Conc
0 0
1
5 0 .0
ENTR
EXIT
SETUP C.3
RANGE CONTROL MENU
MODE
SET UNIT
EXIT
The
LOW
and
HIGH
Ranges have separate
slopes and offsets for
computing the carbon
monoxide concentration.
The two ranges must be
independently calibrated.
Содержание 300E
Страница 40: ...Getting Started Model 300E Carbon Monoxide Analyzer Instruction Manual 40 P N 04288 REV A INTENTIONALLY BLANK ...
Страница 122: ...Operating Instructions Model 300E Carbon Monoxide Analyzer Instruction Manual 122 P N 04288 REV A INTENTIONALLY BLANK ...
Страница 186: ...THEORY OF OPERATION Model 300E Carbon Monoxide Analyzer Instruction Manual 186 P N 04288 Rev A INTENTIONALLY BLANK ...
Страница 224: ...Model 300E Carbon Monoxide Analyzer Instruction Manual Appendix A P N 04288 REV A INTENTIONALLY BLANK ...
Страница 226: ...Model 300E Carbon Monoxide Analyzer Instruction Manual Appendix B P N 04288 REV A INTENTIONALLY BLANK ...
Страница 228: ...Model 300E Carbon Monoxide Analyzer Instruction Manual Appendix C P N 04288 REV A INTENTIONALLY BLANK ...
Страница 230: ...Model 300E Carbon Monoxide Analyzer Instruction Manual Appendix D P N 04288 REV A INTENTIONALLY BLANK ...
Страница 236: ......
Страница 252: ......