Operating Instructions
Model 300E Carbon Monoxide Analyzer
Instruction Manual
74
P/N 04288 REV: A
ANALOG
A1 A2 A3
A4
1 2 3 4 5 6 7 8
Figure 6–2: Analog Output Pin Assignments
6.5.1.
Voltage Output Range Selection and Offset Adjustment
To select an output signal level or add a DC offset to one of the analyzer Analog
Output Channels press:
Calibrates Automatically
Exit to Return
to the main
Sample Display
DIAG
ANALOG I / O CONFIGURATION
PREV
NEXT
ENTR
EXIT
DIAG AIO
AIN A/C FREQUENCY:
60 HZ
SET>
EDIT
EXIT
DIAG AIO
AOUT CALIBRATED:
NO
< SET
SET>
CAL
EXIT
FROM ANALOG I/O CONFIGURATION MENU
Toggle these
Keys to set the
DC Offset
of the selected
channel
Exit to Return
to the I/O
Configuration
Menu
DIAG AIO
CONC_OUT_2 RANGE: 5V
SET> EDIT
EXIT
DIAG AIO
OUTPUT RANGE:
5V
0.1V 1V 5V 10V CURR
ENTR
EXIT
DIAG AIO
CONC_OUT_2 REC OFS: 0 mV
< SET
SET> EDIT
EXIT
DIAG AIO
RECORD OFFSET: 0 MV
+ 0 0 0 0
ENTR
EXIT
DIAG AIO
CONC_OUT_2 CALIBRATED:
NO
<SET
CAL
EXIT
DIAG AIO
CONC_OUT_2 CALIBRATED:
YES
<SET CAL
EXIT
These Keys set
the Signal Level
of the selected
channel
DIAG AIO
CONC_OUT_2:5V, CAL
< SET SET>
EDIT
EXIT
Press
SET>
to select the Analog Output
channel to be configured:
DISPLAYED
AS =
CHANNEL
CONC_OUT_1 =
A1
CONC_OUT_2 =
A2
TEST
OUTPUT =
A3
Then Press
EDIT
to continue
Pressing
ENTR
records the new
setting and returns
to the previous
menu
Pressing
EXIT
ignores the new
setting and returns
to the previous
menu
Содержание 300E
Страница 40: ...Getting Started Model 300E Carbon Monoxide Analyzer Instruction Manual 40 P N 04288 REV A INTENTIONALLY BLANK ...
Страница 122: ...Operating Instructions Model 300E Carbon Monoxide Analyzer Instruction Manual 122 P N 04288 REV A INTENTIONALLY BLANK ...
Страница 186: ...THEORY OF OPERATION Model 300E Carbon Monoxide Analyzer Instruction Manual 186 P N 04288 Rev A INTENTIONALLY BLANK ...
Страница 224: ...Model 300E Carbon Monoxide Analyzer Instruction Manual Appendix A P N 04288 REV A INTENTIONALLY BLANK ...
Страница 226: ...Model 300E Carbon Monoxide Analyzer Instruction Manual Appendix B P N 04288 REV A INTENTIONALLY BLANK ...
Страница 228: ...Model 300E Carbon Monoxide Analyzer Instruction Manual Appendix C P N 04288 REV A INTENTIONALLY BLANK ...
Страница 230: ...Model 300E Carbon Monoxide Analyzer Instruction Manual Appendix D P N 04288 REV A INTENTIONALLY BLANK ...
Страница 236: ......
Страница 252: ......