Operating Basics
The ALARM menu contains the following selections:
ALARM STATUS:
Displays the alarm status detected inside the instrument.
The following status messages may be displayed.
No Alarm:
No errors have been detected.
Genlock: Near Loss of Lock:
The frequency of the genlock signal and
the internal reference signal have shifted considerably, and the TG700 is
near loss of lock. If this message appears, check the frequency of the
genlock signal or the internal reference signal of the TG700.
Genlock: Loss of Lock:
The TG700 cannot lock to the genlock signal.
Genlock: No Signal:
A genlock signal is not connected.
Fan Fault:
The fan on the rear panel has stopped working.
NOTE.
In the alarm status display, the instrument displays an error message
that has been detected in the past. In order to refresh the display, you need
to access another menu item (press the CANCEL button), and then reselect the
ALARM STATUS menu item.
ALARM RESET:
Resets the alarm status. The TG700 holds the alarm status
even if the cause of the alarm is eliminated. Use this menu item when you
need to reset the status. If the cause of the error is not eliminated, the alarm
status is set again when you reset the status.
GPI ALARM OUTPUT:
Enables or disables the alarm signal output. Use
the left (
◄
) or right (
►
) arrow button to select between
Enable
and
Disable
,
and then press the
ENTER
button to con
fi
rm the selection. When the GPI
alarm output is enabled, an alarm signal is output from pin 6 of the 10
BASE-T connector when any error is detected inside the instrument. Since
the output signal is an active low; the signal level for pin 6 goes from High to
Low when any errors occur.
GPI ALARM SELECT:
Selects which GPI alarms are enabled or disabled.
Select this menu item, and then press the
ENTER
button to access the GPI
ALARM SELECT submenu. (See page 2-39.)
AGL7 ALARM OUTPUT:
Enables or disables the alarm output from one
of the BLACK outputs on the AGL7 module. Use the left (
◄
) or right (
►
)
arrow button to select
Disable
,
BLACK 1
,
BLACK 2
, or
BLACK 3
. When
you select BLACK 1, BLACK 2, or BLACK 3 and an enabled error occurs,
the output signal from the selected connector turns off for about 60 ms. When
2–38
TG700 TV Signal Generator Platform User Manual
Содержание TG700
Страница 1: ...xx TG700 TV Signal Generator Platform ZZZ User Manual P071197003 071 1970 03...
Страница 2: ......
Страница 27: ...Getting Started...
Страница 28: ......
Страница 34: ...Getting Started Figure 1 1 Installing the rackmount hardware 1 6 TG700 TV Signal Generator Platform User Manual...
Страница 67: ...Operating Basics...
Страница 68: ......
Страница 111: ...Modules...
Страница 112: ......
Страница 337: ...Index...
Страница 338: ......