Appendix A: Warranted Speci
fi
cations
Table 1: RF Channel Characteristics (cont.)
Characteristic
Description
√
Spurious Response (SFDR)
2
nd
harmonic distortion: >100 MHz, < –55 dBc
2nd harmonic distortion: > 100 MHz, < –60 dBc (typical)
With auto settings on and signals 10 dB below reference level
2
nd
harmonic distortion:9 kHz to 100 MHz, < –55 dBc
2
nd
harmonic distortion: 9 kHz to 100 MHz, < –60 dBc (typical)
With auto settings on and signals 10 dB below reference level and reference level
≤
–15 dBm
3
rd
order harmonic distortion: >100 MHz < –53 dBc
3
rd
order harmonic distortion: >100 MHz < –58 dBc (typical)
With auto settings on and signals 10 dB below reference level
3
rd
order harmonic distortion: 9 kHz to 100 MHz, < –55 dBc
3
rd
order harmonic distortion: 9 kHz to 100 MHz, < –60 dBc (typical)
With auto settings on and signals 10 dB below reference level and reference level
≤
–15 dBm
2
nd
order intermodulation distortion: >100 Mhz, < –55 dBc
2
nd
order intermodulation distortion: >100 MHz, < –60 dBc (typical)
With auto settings on and signals 10 dB below reference level
2
nd
order intermodulation distortion: 9 kHz to 15 MHz, < –55 dBc
2
nd
order intermodulation distortion: 9 kHz to 15 MHz, < –60 dBc (typical)
With auto settings on and signals 10 dB below reference level and reference level
≤
–15 dBm
3
rd
order intermodulation distortion: >15 MHz, < –55 dBc
3
rd
order intermodulation distortion: >15 MHz < –60 dBc (typical)
With auto settings on and signals 10 dB below reference level
3
rd
order intermodulation distortion:9 kHz to 15 MHz, < –55 dBc
3
rd
order intermodulation distortion: 9 kHz to 15 MHz, < –60 dBc (typical)
With auto settings on and signals 10 dB below reference level and reference level
≤
–15 dBm
–45 dBc (–50 dBc, typical) for side bands < 25 kHz offset from the carrier
√
Residual Spurious Response
< –78 dBm
< –67 dBm at 2.5 GHz
< –76 dBm at 1.25 GHz
≤
–15 dBm reference level and RF input terminated with 50
Ω
.
√
Reference Frequency Error
(Cumulative)
Cumulative Error: ±10 x 10
-6
Includes allowances for Aging per Year, Reference Frequency Calibration Accuracy,
and Temperature Stability.
Valid over the recommended 1 year calibration interval, from –10 °C to +55 °C .
196
MDO3000 Series Oscilloscopes User Manual
Содержание MDO3012
Страница 1: ...x MDO3000 Series Mixed Domain Oscilloscopes ZZZ User Manual P077096802 077 0968 02...
Страница 2: ......
Страница 3: ...MDO3000 Series Mixed Domain Oscilloscopes ZZZ User Manual xx www tek com 077 0968 02...
Страница 10: ...Table of Contents iv MDO3000 Series Oscilloscopes User Manual...
Страница 236: ...Appendix D OpenSSL License 214 MDO3000 Series Oscilloscopes User Manual...