Installation
Optional application module packages extend the capability of your oscilloscope.
You can physically install up to two application modules at one time. Application modules can go into the slots with a window
in the upper right corner of the front panel. One additional slot is directly behind the one that you can see. To use the hidden
slot, install the module with the label facing away from you.
Some of the modules have licenses that allow you to transfer the license between your application modules and the
oscilloscope. You can keep each license in the module, which will allow you to move the module from one instrument to
another. Alternatively, you can move the license from the module to the oscilloscope. This approach will allow you to store
the module separately from the oscilloscope for safe keeping. This approach will also allow you to use more than two
applications on your oscilloscope simultaneously. To transfer a license from a module to your oscilloscope or from your
oscilloscope to a module:
1.
Turn off the power to the oscilloscope. Insert the application modules into the oscilloscope. Turn on the power.
2.
Push
Utility
on the front panel. If needed, push
Utility Page
on the lower menu and turn the
Multipurpose a
knob to
select
Con
fi
g
. Push
Manage Modules and Options
on the lower menu, and then push
License Type
on the side menu
until “Modules” is selected. The licenses contained in the oscilloscope will be listed in the side menu. Push the button
next to the appropriate license to transfer. You may transfer up to two licenses at one time.
3.
After you turn off the power to the oscilloscope, you can remove the physical application module from the oscilloscope.
Refer to the
MDO3000 Series Oscilloscopes Application Module Installation Manual
that came with your application module
for instructions on installing and testing an application module.
NOTE.
If you transfer a license from a module to an oscilloscope, the module will not work on another oscilloscope until
you transfer the license back from the oscilloscope to the module. Consider putting the physical module in an envelope
or other storage with a label recording the date, module name, model and serial number of the oscilloscope which holds
the license. This will help prevent confusion later if someone
fi
nds the module, installs it in some other oscilloscope,
and wonders why it does not work.
Upgrading Bandwidth
You may increase the bandwidth of an instrument when your project requirements demand higher performance. Do this
by purchasing an upgrade.
1 GHz upgrades require Tek Service installation and option IFC (calibration).
Model to be upgraded
Bandwidth before
upgrade
Bandwidth after
upgrade
Order this product
MDO3012
100 MHz
200 MHz
MDO3BW1T22
100 MHz
350 MHz
MDO3BW1T32
100 MHz
500 MHz
MDO3BW1T52
100 MHz
1 GHz
MDO3BW1T102
200 MHz
350 MHz
MDO3BW2T32
200 MHz
500 MHz
MDO3BW2T52
MDO3000 Series Oscilloscopes User Manual
15
Содержание MDO3012
Страница 1: ...x MDO3000 Series Mixed Domain Oscilloscopes ZZZ User Manual P077096802 077 0968 02...
Страница 2: ......
Страница 3: ...MDO3000 Series Mixed Domain Oscilloscopes ZZZ User Manual xx www tek com 077 0968 02...
Страница 10: ...Table of Contents iv MDO3000 Series Oscilloscopes User Manual...
Страница 236: ...Appendix D OpenSSL License 214 MDO3000 Series Oscilloscopes User Manual...