Analyze Waveform or Trace Data
Automatic Peak Markers
Automatic peak markers are on by default and assist with quickly identifying the frequency and amplitude of peaks in
the spectrum.
1.
The Reference Marker is placed on the
highest amplitude peak. It is marked with
a red R in a triangle.
2.
The automatic markers indicate
frequency and amplitude.
3.
Absolute readouts show the actual
frequency and amplitude of the
automatic markers.
4.
Delta readouts show the frequency and
amplitude of the automatic markers
relative to the reference marker.
In the screen shot shown below, a marker has been placed on each of the obvious peaks in the display. The Reference
Marker is the highest peak. It is marked with the red R in a triangle, and its readout is shown in red text.
1.
Reference marker
2.
Automatic markers
Use
Threshold
and
Excursion
to de
fi
ne which peaks are marked.
The threshold is a minimum amplitude that a signal must cross to be a valid peak. If the threshold is lower, more peaks will
tend to qualify to have markers. If the threshold is higher, fewer peaks tend to qualify to have markers.
The excursion is how far a signal needs to fall in amplitude between marked peaks to be another valid peak. If the excursion
is low, more peaks will tend to qualify to have associated markers. If the excursion is high, fewer peaks will tend to qualify to
have associated markers.
122
MDO3000 Series Oscilloscopes User Manual
Содержание MDO3012
Страница 1: ...x MDO3000 Series Mixed Domain Oscilloscopes ZZZ User Manual P077096802 077 0968 02...
Страница 2: ......
Страница 3: ...MDO3000 Series Mixed Domain Oscilloscopes ZZZ User Manual xx www tek com 077 0968 02...
Страница 10: ...Table of Contents iv MDO3000 Series Oscilloscopes User Manual...
Страница 236: ...Appendix D OpenSSL License 214 MDO3000 Series Oscilloscopes User Manual...