Acquire the Signal
LIN Bus
To acquire data from a LIN bus, you need to also set up these items:
1.
If you selected
LIN
, push
De
fi
ne Inputs
and
the appropriate side menu choices.
Bus B1
LIN
De
fi
ne
Inputs
Thresholds
Con
fi
gure
B1 Label
LIN
Bus
Display
Event
Table
2.
Turn
Multipurpose a
to select the channel
connected to the LIN bus source.
LIN Input
(a) 1
3.
Turn
Multipurpose a
to set the
Sample
Point
from 5% to 95% of the position within
the bit period or the unit interval.
Sample
Point
50%
4.
Select the
Polarity
to match the LIN bus
being acquired.
Polarity
Normal
(High=1)
Polarity
Inverted
(High=0)
5.
Push
Con
fi
gure
and the appropriate side
menu choices.
Bus B1
LIN
De
fi
ne
Inputs
Thresholds
Con
fi
gure
B1 Label
LIN
Bus
Display
Event
Table
6.
Push
Bit Rate
, and turn
Multipurpose a
to
select from the list of prede
fi
ned bit rates.
Alternatively, you can set the bit rate to a
speci
fi
c value. To do so, select
Custom
, and
then turn
Multipurpose b
to set the bit rate
from 800 bps to 100,000 bps.
Bit Rate
(a)
19.2K bps
7.
Push
LIN Standard
, and turn
Multipurpose
a
to select the appropriate standard.
LIN
Standard
v1.x
8.
Push
Include Parity Bits with Id
to select
whether or not to include parity bits.
Include
Parity Bits
with Id
On
Off
MDO3000 Series Oscilloscopes User Manual
77
Содержание MDO3012
Страница 1: ...x MDO3000 Series Mixed Domain Oscilloscopes ZZZ User Manual P077096802 077 0968 02...
Страница 2: ......
Страница 3: ...MDO3000 Series Mixed Domain Oscilloscopes ZZZ User Manual xx www tek com 077 0968 02...
Страница 10: ...Table of Contents iv MDO3000 Series Oscilloscopes User Manual...
Страница 236: ...Appendix D OpenSSL License 214 MDO3000 Series Oscilloscopes User Manual...